MR-MOTUS: model-based non-rigid motion estimation for MR-guided radiotherapy using a reference image and minimal k-space data

被引:35
作者
Huttinga, Niek R. F. [1 ]
van den Berg, Cornelis A. T. [1 ]
Luijten, Peter R. [1 ]
Sbrizzi, Alessandro [1 ]
机构
[1] Univ Med Ctr, Imaging Div, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
关键词
MR-guided radiotherapy; non-rigid motion estimation; inverse problems; model-based reconstruction; high frame rate motion; NAVIGATOR ECHOES; RECONSTRUCTION; REGISTRATION; THERAPIES; FRAMEWORK; TRACKING; COIL;
D O I
10.1088/1361-6560/ab554a
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Time-resolved motion estimation from MRI data has received an increasing amount of interest due to the advent of the MR-Linac. The combination of an MRI scanner and a linear accelerator enables radiation plan adaptation based on internal organ motion estimated from MRI data. However, time-resolved estimation of this motion from MRI data still remains a challenge. In light of this application, we propose MR-MOTUS, a framework to estimate non-rigid 3D motion from minimal k-space data. MR-MOTUS consists of two main components: (1) a signal model that explicitly relates the k-space signal of a deforming object to non-rigid motion-fields and a reference image, and (2) model-based reconstructions of the non-rigid motion-fields directly from k-space data. Using an a priori available reference image and the fact that internal body motion exhibits a high level of spatial correlation, we represent the motion-fields in a low-dimensional space and reconstruct them from minimal k-space data that can be acquired very rapidly. The signal model is validated through numerical experiments with a digital 3D phantom and motion-fields are reconstructed from retrospectively undersampled in vivo head and abdomen data using various undersampling strategies. A comparison is made with state-of-the-art image registration performed on images reconstructed from the same undersampled data. Results show that MR-MOTUS reconstructs in vivo 3D rigid head motion from 474-fold retrospectively downsampled k-space data, and in vivo non-rigid 3D respiratory motion from 63-fold retrospectively undersampled k-space data. Preliminary results on prospectively undersampled data acquired with a 2D golden angle acquisition during free-breathing demonstrate the practical feasibility of the method.
引用
收藏
页数:19
相关论文
共 44 条
  • [31] Nonrigid registration using free-form deformations: Application to breast MR images
    Rueckert, D
    Sonoda, LI
    Hayes, C
    Hill, DLG
    Leach, MO
    Hawkes, DJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (08) : 712 - 721
  • [32] Sbrizzi A, 2019, P INT SOC MAG RESON, V27, P795
  • [33] Motion prediction in MRI-guided radiotherapy based on interleaved orthogonal cine-MRI
    Seregni, M.
    Paganelli, C.
    Lee, D.
    Greer, P. B.
    Baroni, G.
    Keall, P. J.
    Riboldi, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (02) : 872 - 887
  • [34] Navigators for motion detection during real-time MRI-guided radiotherapy
    Stam, Mette K.
    Crijns, Sjoerd P. M.
    Zonnenberg, Bernard A.
    Barendrecht, Maurits M.
    van Vulpen, Marco
    Lagendijk, Jan J. W.
    Raaymakers, Bas W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) : 6797 - 6805
  • [35] Image-driven, model-based 3D abdominal motion estimation for MR-guided radiotherapy
    Stemkens, Bjorn
    Tijssen, Rob H. N.
    de Senneville, Baudouin Denis
    Lagendijk, Jan J. W.
    van den Berg, Cornelis A. T.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14) : 5335 - 5355
  • [36] 4D tumor centroid tracking using orthogonal 2D dynamic MRI: Implications for radiotherapy planning
    Tryggestad, Erik
    Flammang, Aaron
    Hales, Russell
    Herman, Joseph
    Lee, Junghoon
    McNutt, Todd
    Roland, Teboh
    Shea, Steven M.
    Wong, John
    [J]. MEDICAL PHYSICS, 2013, 40 (09)
  • [37] Uecker M, 2019, MRIRECON BART VERSIO, DOI 10.5281/zenodo.3376744
  • [38] Motion corrected compressed sensing for free-breathing dynamic cardiac MRI
    Usman, Muhammad
    Atkinson, David
    Odille, Freddy
    Kolbitsch, Christoph
    Vaillant, Ghislain
    Schaeffter, Tobias
    Batchelor, Philip G.
    Prieto, Claudia
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (02) : 504 - 516
  • [39] Real-time rigid body motion correction and shimming using cloverleaf navigators
    van der Kouwe, Andre J. W.
    Benner, Thomas
    Dale, Anders M.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (05) : 1019 - 1032
  • [40] Spherical navigator echoes for full 3D rigid body motion measurement in MRI
    Welch, EB
    Manduca, A
    Grimm, RC
    Ward, HA
    Jack, CR