Toward a more realistic prediction of peri-prosthetic micromotions

被引:20
作者
van der Ploeg, Bas [1 ]
Tarala, Maria [2 ]
Homminga, Jasper [1 ]
Janssen, Dennis [2 ]
Buma, Pieter [2 ]
Verdonschot, Nico [1 ,2 ]
机构
[1] Univ Twente, Lab Biomech Engn, NL-7500 AE Enschede, Netherlands
[2] Radboud Univ Nijmegen, Orthopaed Res Lab 357, Med Ctr, NL-6500 HB Nijmegen, Netherlands
关键词
total hip arthroplasty; finite element analysis; micromotion; musculoskeletal modeling; PRIMARY STABILITY; HIP STEM; IMPLANT MICROMOTION; IN-VIVO; INTERFACE; FIXATION; LOAD;
D O I
10.1002/jor.22041
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The finite element (FE) method has become a common tool to evaluate peri-prosthetic micromotions in cementless total hip arthroplasty. Often, only the peak joint load and a selected number of muscle loads are applied to determine micromotions. Furthermore, the applied external constraints are simplified (diaphyseal fixation), resulting in a non-physiological situation. In this study, a scaled musculoskeletal model was used to extract a full set of muscle and hip joint loads occurring during a walking cycle. These loads were applied incrementally to an FE model to analyze micromotions. The relation between micromotions and external loads was investigated, and how micromotions during a full loading cycle compared to those calculated when applying a peak load only. Finally, the effect of external constraints was analyzed (full model vs. diaphyseal fixation and reduced number of muscle loads). Relatively large micromotions were found during the swing phase when the hip joint forces were relatively low. Maximal micromotions, however, did concur with the peak hip joint force. Applying only a peak joint force resulted in peak micromotions similar to those found when full walking cycle loads were applied. The magnitude and direction of the micromotions depended on the applied muscle loads, but not on external constraints. (C) 2012 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 30:11471154, 2012
引用
收藏
页码:1147 / 1154
页数:8
相关论文
共 27 条
  • [1] Finite element modelling of primary hip stem stability: The effect of interference fit
    Abdul-Kadir, Mohammed Rafiq
    Hansen, Ulrich
    Klabunde, Ralf
    Lucas, Duncan
    Amis, Andrew
    [J]. JOURNAL OF BIOMECHANICS, 2008, 41 (03) : 587 - 594
  • [2] Baleani M, 2000, J BIOMED MATER RES, V50, P605, DOI 10.1002/(SICI)1097-4636(20000615)50:4<605::AID-JBM17>3.0.CO
  • [3] 2-P
  • [4] The effect of muscle loading on the simulation of bone remodelling in the proximal femur
    Bitsakos, C
    Kerner, J
    Fisher, I
    Amis, AA
    [J]. JOURNAL OF BIOMECHANICS, 2005, 38 (01) : 133 - 139
  • [5] Differences in stiffness of the interface between a cementless porous implant and cancellous bone in vivo in dogs due to varying amounts of implant motion
    Bragdon, CR
    Burke, D
    Lowenstein, JD
    OConnor, DO
    Ramamurti, B
    Jasty, M
    Harris, WH
    [J]. JOURNAL OF ARTHROPLASTY, 1996, 11 (08) : 945 - 951
  • [6] Three-dimensional primary stability of cementless femoral stems
    Buhler, DW
    Berlemann, U
    Lippuner, K
    Jaeger, P
    Nolte, LP
    [J]. CLINICAL BIOMECHANICS, 1997, 12 (02) : 75 - 86
  • [7] Burke DW., 1991, T ORTHOP RES SOC, V16, P103
  • [8] Influence of muscle forces on femoral strain distribution
    Duda, GN
    Heller, M
    Albinger, J
    Schulz, O
    Schneider, E
    Claes, L
    [J]. JOURNAL OF BIOMECHANICS, 1998, 31 (09) : 841 - 846
  • [9] ENGH CA, 1992, CLIN ORTHOP RELAT R, P13
  • [10] Determination of muscle loading at the hip joint for use in pre-clinical testing
    Heller, MO
    Bergmann, G
    Kassi, JP
    Claes, L
    Haas, NP
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2005, 38 (05) : 1155 - 1163