Single Cell Microwave Biosensor for Monitoring Cellular Response to Electrochemotherapy

被引:10
|
作者
Tamra, Amar [1 ,2 ,3 ]
Zedek, Amel [1 ,4 ]
Rols, Marie-Pierre [2 ]
Dubuc, David [1 ]
Grenier, Katia [1 ]
机构
[1] Natl Sci Res Ctr, Lab Anal & Architecture Syst, F-31031 Toulouse, France
[2] Inst Pharmacol & Struct Biol, F-31077 Toulouse, France
[3] King Abdullah Univ Sci & Technol, Thuwal, Saudi Arabia
[4] Landways Management, London, England
关键词
Microwave measurement; Spectroscopy; Cells (biology); Microwave theory and techniques; Biology; Dielectric measurement; Frequency measurement; Biosensor; electrochemotherapy; microwave dielectric spectroscopy; single Cell; THP-1; cells; ON-CHIP ELECTROPORATION; IMPEDANCE SPECTROSCOPY; BLEOMYCIN; ELECTROPERMEABILIZATION; PERSPECTIVES; DELIVERY; ARRAYS; DRUG;
D O I
10.1109/TBME.2022.3170267
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a 40 GHz microwave biosensor used to monitor and characterize single cells (THP-1) subjected to electrochemotherapy and obtain an electronic signature of the treatment efficiency. This biosensor proposes a non-destructive and label-free technique that first allows, with the rapid measurement of single untreated cells in their culture medium, the extraction of two frequency-dependent dielectric parameters, the capacitance (C (f)) and the conductance (G (f)). Second, this technique can powerfully reveal the effects of a chemical membrane permeabilizing treatment (Saponin). At last, it permits us to detect, and predict, the potentiation of a molecule classically used in chemotherapy (Bleomycin) when combined with the application of electric pulses (principle of electrochemotherapy). Treatment-affected cells show a decrease in the capacitive and conductive contrasts, indicating damages at the cellular levels. Along with these results, classical biological tests are conducted. Statistical analysis points out a high correlation rate (R-2>0.97), which clearly reveals the reliability and efficacy of our technique and makes it an attractive technique for biology related researches and personalized medicine.
引用
收藏
页码:3407 / 3414
页数:8
相关论文
共 50 条
  • [1] Microwave Monitoring of Single Cell Monocytes Subjected to Electroporation
    Tamra, Amar
    Dubuc, David
    Rols, Marie-Pierre
    Grenier, Katia
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2017, 65 (09) : 3512 - 3518
  • [2] Extension to In Situ Single-Cell Electroporation of a Microwave Biosensor
    Calvel, Anne
    Peytral-Rieu, Olivia
    Rols, Marie-Pierre
    Dubuc, David
    Grenier, Katia
    IEEE MICROWAVE AND WIRELESS TECHNOLOGY LETTERS, 2024, 34 (06): : 849 - 852
  • [3] Single-cell bioelectrical impedance platform for monitoring cellular response to drug treatment
    Asphahani, Fareid
    Wang, Kui
    Thein, Myo
    Veiseh, Omid
    Yung, Sandy
    Xu, Jian
    Zhang, Miqin
    PHYSICAL BIOLOGY, 2011, 8 (01)
  • [4] Microwave biosensor dedicated to the dielectric spectroscopy of a single alive biological cell in its culture medium
    Chen, T.
    Artis, F.
    Dubuc, D.
    Fournie, J-J.
    Poupot, M.
    Grenier, K.
    2013 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM DIGEST (IMS), 2013,
  • [5] High reliability Microfluidic biosensor for single cell impedance cytometry
    Claudel, J.
    Nadi, M.
    El Mazria, O.
    Kourtiche, D.
    2017 ELEVENTH INTERNATIONAL CONFERENCE ON SENSING TECHNOLOGY (ICST), 2017, : 1 - 5
  • [6] Insights into the HyPer biosensor as molecular tool for monitoring cellular antioxidant capacity
    Hernandez, Helen
    Parra, Alejandra
    Tobar, Nicolas
    Molina, Jessica
    Kallens, Violeta
    Hidalgo, Miltha
    Varela, Diego
    Martinez, Jorge
    Porras, Omar
    REDOX BIOLOGY, 2018, 16 : 199 - 208
  • [7] Microwave dielectric spectroscopy for single cell irreversible electroporation monitoring
    Tamra, Amar
    Deburghgraeve, Marie
    Dubuc, David
    Rols, Marie-Pierre
    Grenier, Katia
    2016 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2016,
  • [8] An optical detection method for analyzing the cellular response to paclitaxel at the single cell level
    Wang, Yijia
    Zhang, Yuquan
    Zhang, Shiwu
    Zhao, Zhenying
    Min, Changjun
    Liu, Jun
    Yuan, X. -C.
    ANALYTICAL METHODS, 2016, 8 (18) : 3698 - 3703
  • [9] The single-cell biosensor for monitoring extracellular action potential and evaluating drugs
    Wang, Ping
    Liu, Qingjun
    Xu, Ying
    Cai, Hua
    PROGRESS ON POST-GENOME TECHNOLOGIES, 2006, : 350 - 354
  • [10] BiCMOS Microfluidic Sensor for Single Cell Label-Free Monitoring Through Microwave Intermodulation
    Palego, C.
    Perry, G.
    Hancock, C.
    Hjeij, F.
    Dalmay, C.
    Bessandou, A.
    Blondy, P.
    Pothier, A.
    Lalloue, F.
    Bessette, B.
    Begaud, G.
    Jauberteau, M-O.
    Kaynak, C. B.
    Wietstruck, M.
    Kaynak, M.
    Casbon, M.
    Benedikt, J.
    Barrow, D.
    Porch, A.
    2016 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2016,