A Preliminary Investigation of the Air-Bone Gap: Changes in Intracochlear Sound Pressure With Air- and Bone-conducted Stimuli After Cochlear Implantation

被引:29
作者
Hartl, Renee M. Banakis [1 ]
Mattingly, Jameson K. [1 ]
Greene, Nathaniel T. [1 ,2 ]
Jenkins, Herman A. [1 ]
Cass, Stephen P. [1 ]
Tollin, Daniel J. [1 ,2 ]
机构
[1] Univ Colorado, Dept Otolaryngol, Sch Med, 12631 E 17th Ave,MS B205, Aurora, CO 80045 USA
[2] Univ Colorado, Dept Physiol & Biophys, Sch Med, Aurora, CO USA
关键词
Air-bone gap; Cochlear implant; Electroacoustic stimulation; Hearing preservation; Intracochlear pressures; MIDDLE-EAR IMPLANT; ELECTRIC-ACOUSTIC STIMULATION; HEARING-LOSS; RESIDUAL HEARING; INNER-EAR; TEMPORAL BONES; FIBROUS TISSUE; PRESERVATION; CONSERVATION; RECOGNITION;
D O I
10.1097/MAO.0000000000001184
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis:A cochlear implant electrode within the cochlea contributes to the air-bone gap (ABG) component of postoperative changes in residual hearing after electrode insertion.Background:Preservation of residual hearing after cochlear implantation has gained importance as simultaneous electric-acoustic stimulation allows for improved speech outcomes. Postoperative loss of residual hearing has previously been attributed to sensorineural changes; however, presence of increased postoperative ABG remains unexplained and could result in part from altered cochlear mechanics. Here, we sought to investigate changes to these mechanics via intracochlear pressure measurements before and after electrode implantation to quantify the contribution to postoperative ABG.Methods:Human cadaveric heads were implanted with titanium fixtures for bone conduction transducers. Velocities of stapes capitulum and cochlear promontory between the two windows were measured using single-axis laser Doppler vibrometry and fiber-optic sensors measured intracochlear pressures in scala vestibuli and tympani for air- and bone-conducted stimuli before and after cochlear implant electrode insertion through the round window.Results:Intracochlear pressures revealed only slightly reduced responses to air-conducted stimuli consistent with previous literature. No significant changes were noted to bone-conducted stimuli after implantation. Velocities of the stapes capitulum and the cochlear promontory to both stimuli were stable after electrode placement.Conclusion:Presence of a cochlear implant electrode causes alterations in intracochlear sound pressure levels to air, but not bone, conducted stimuli and helps to explain changes in residual hearing noted clinically. These results suggest the possibility of a cochlear conductive component to postoperative changes in hearing sensitivity.
引用
收藏
页码:1291 / 1299
页数:9
相关论文
共 46 条
  • [1] Development and evaluation of an improved cochlear implant electrode design for electric acoustic stimulation
    Adunka, O
    Kiefer, J
    Unkelbach, MH
    Lehnert, T
    Gstoettner, W
    [J]. LARYNGOSCOPE, 2004, 114 (07) : 1237 - 1241
  • [2] Conservation of residual acoustic hearing after cochlear implantation
    Balkany, Thomas J.
    Connell, Sarah S.
    Hodges, Annelle V.
    Payne, Stacy L.
    Telischi, Fred F.
    Eshraghi, Adrien A.
    Angeli, Simon I.
    Germani, Ross
    Messiah, Sarah
    Arheart, Kristopher L.
    [J]. OTOLOGY & NEUROTOLOGY, 2006, 27 (08) : 1083 - 1088
  • [3] Residual Hearing Preservation After Pediatric Cochlear Implantation
    Brown, Ryan F.
    Hullar, Timothy E.
    Cadieux, Jamie H.
    Chole, Richard A.
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (08) : 1221 - 1226
  • [4] Predicting the effect of post-implant cochlear fibrosis on residual hearing
    Choi, CH
    Oghalai, JS
    [J]. HEARING RESEARCH, 2005, 205 (1-2) : 193 - 200
  • [5] Conductive Component After Cochlear Implantation in Patients With Residual Hearing Conservation
    Chole, Richard A.
    Hullar, Timothy E.
    Potts, Lisa G.
    [J]. AMERICAN JOURNAL OF AUDIOLOGY, 2014, 23 (04) : 359 - 364
  • [6] Techniques to Improve the Efficiency of a Middle Ear Implant: Effect of Different Methods of Coupling to the Ossicular Chain
    Deveze, Arnaud
    Koka, Kanthaiah
    Tringali, Stephane
    Jenkins, Herman A.
    Tollin, Daniel J.
    [J]. OTOLOGY & NEUROTOLOGY, 2013, 34 (01) : 158 - 166
  • [7] Active Middle Ear Implant Application in Case of Stapes Fixation: A Temporal Bone Study
    Deveze, Arnaud
    Koka, Kanthaiah
    Tringali, Stephane
    Jenkins, Herman A.
    Tollin, Daniel J.
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (07) : 1027 - 1034
  • [8] Effect of cochlear implant electrode insertion on middle-ear function as measured by intra-operative laser Doppler vibrometry
    Donnelly, N.
    Bibas, A.
    Jiang, D.
    Bamiou, D-E
    Santulli, C.
    Jeronimidis, G.
    O'Connor, A. Fitzgerald
    [J]. JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2009, 123 (07) : 723 - 729
  • [9] The benefits of combining acoustic and electric stimulation for the recognition of speech, voice and melodies
    Dorman, Michael F.
    Gifford, Rene H.
    Spahr, Anthony J.
    McKarns, Sharon A.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2008, 13 (02) : 105 - 112
  • [10] Pattern of hearing loss in a rat model of cochlear implantation trauma
    Fayad, JN