Ultra-trace arsenic and mercury speciation and determination in blood samples by ionic liquid-based dispersive liquid-liquid microextraction combined with flow injection-hydride generation/cold vapor atomic absorption spectroscopy

被引:25
作者
Shirkhanloo, Hamid [1 ]
Khaligh, Aisan [2 ]
Mousavi, Hassan Zavvar [2 ]
Eskandari, Mohammad Mehdi [3 ]
Miran-Beigi, Ali Akbar [3 ]
机构
[1] IPIHRI, IPIOEHRC, Tehran 1485733111, Iran
[2] Semnan Univ, Dept Chem, Semnan 3513119111, Iran
[3] RIPI, Tehran 1485733111, Iran
关键词
arsenic; mercury; microwave assisted-UV system; ionic liquid-based dispersive liquid-liquid microextraction; hydride generation/cold vapor atomic absorption spectrometry; HPLC-ICP-MS; BIOLOGICAL SAMPLES; WATER SAMPLES; HUMAN URINE; SPECTROMETRY; PRECONCENTRATION; METHYLMERCURY; EXTRACTION; DIGESTION; EXPOSURE;
D O I
10.1515/chempap-2015-0086
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A simple, fast, and sensitive method for speciation and determination of As (III, V) and Hg (II, R) in human blood samples based on ionic liquid-dispersive liquid-liquid microextraction (IL-DLLME) and flow injection hydride generation/cold vapor atomic absorption spectrometry (FI-HG/CV-AAS) has been developed. Tetraethylthiuram disulfide, mixed ionic liquids (hydrophobic and hydrophilic ILs) and acetone were used in the DLLME step as the chelating agent, extraction and dispersive solvents, respectively. Using a microwave assisted-UV system, organic mercury (R-Hg) was converted to Hg(II) and total mercury amount was measured in blood samples by the presented method. Total arsenic content was determined by reducing As(V) to As(III) with potassium iodide and ascorbic acid in a hydrochloric acid solution. Finally, As(V) and R-Hg were determined by mathematically subtracting the As(III) and Hg(II) content from the total arsenic and mercury, respectively. Under optimum conditions, linear range and detection limit (3 sigma) of 0.1-5.0 mu g L-1 and 0.02 mu g L-1 for As(III) and 0.15-8.50 mu g L-1 and 0.03 mu g L-1 for Hg(II) were achieved, respectively, at low RSD values of < 4 % (N = 10). The developed method was successfully applied to determine the ultra-trace amounts of arsenic and mercury species in blood samples; the validation of the method was performed using standard reference materials. (C) 2015 Institute of Chemistry, Slovak Academy of Sciences
引用
收藏
页码:779 / 790
页数:12
相关论文
共 49 条
[1]   DECOMPOSITION AND STABILITY STUDIES OF METHYLMERCURY IN WATER USING COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
AHMED, R ;
STOEPPLER, M .
ANALYST, 1986, 111 (12) :1371-1374
[2]   Determination of very low levels of dissolved mercury(II) and methylmercury in river waters by continuous flow with on-line UV decomposition and cold-vapor atomic fluorescence spectrometry after pre-concentration on a silica gel-2-mercaptobenzimidazol sorbent [J].
Bagheri, H ;
Gholami, A .
TALANTA, 2001, 55 (06) :1141-1150
[3]   Determination of arsenic in biological fluids by electrothermal atomic absorption spectrometry [J].
Campillo, N ;
Viñas, P ;
López-García, I ;
Hernández-Córdoba, M .
ANALYST, 2000, 125 (02) :313-316
[4]   Advanced oxidation processes for degradation of organomercurials: determination of inorganic and total mercury in urine by FI-CV-AAS [J].
Capelo, JL ;
Maduro, C ;
Mota, AM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (03) :414-416
[5]  
Chen YC, 2002, CANCER EPIDEM BIOMAR, V11, P1427
[6]   Human exposure to mercury: The three modern dilemmas [J].
Clarkson, TW ;
Magos, L ;
Myers, GJ .
JOURNAL OF TRACE ELEMENTS IN EXPERIMENTAL MEDICINE, 2003, 16 (04) :321-343
[7]   A novel ionic liquid/micro-volume back extraction procedure combined with flame atomic absorption spectrometry for determination of trace nickel in samples of nutritional interest [J].
Dadfarnia, Shayesteh ;
Shabani, Ali Mohammad Haji ;
Bidabadi, Mahboubeh Shirani ;
Jafari, Abbas Ali .
JOURNAL OF HAZARDOUS MATERIALS, 2010, 173 (1-3) :534-538
[8]   Analysis of volatile arsenic compounds formed by intestinal microorganisms: rapid identification of new metabolic products by use of simultaneous EI-MS and ICP-MS detection after gas chromatographic separation [J].
Diaz-Bone, Roland A. ;
Hollmann, Markus ;
Wuerfel, Oliver ;
Pieper, Dominik .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (06) :808-814
[9]  
Didi M.A., 2013, AM J ANAL CHEM, P40, DOI [10.4236/ajac.2013.47A006, DOI 10.4236/AJAC.2013.47A006]
[10]   Speciation of inorganic arsenic in alimentary and environmental aqueous samples by using derivative anodic stripping chronopotentiometry (dASCP) [J].
Dugo, G ;
La Pera, L ;
Lo Turco, V ;
Di Bella, G .
CHEMOSPHERE, 2005, 61 (08) :1093-1101