A new navigational apparatus for fixation of acetabular posterior column fractures with percutaneous retrograde lagscrew Design and application

被引:14
|
作者
Zhang, Pijun [1 ]
Tang, Jie [1 ]
Dong, Yonghui [2 ]
Lu, Lu [1 ]
Wang, Shengjie [2 ]
Song, Shifeng [1 ]
Wang, Gang [3 ]
机构
[1] Hainan Med Univ, Dept Orthoped, Affiliated Hosp 2, Haikou, Hainan, Peoples R China
[2] Henan Prov Peoples Hosp, Dept Orthoped, Zhengzhou, Henan, Peoples R China
[3] Nnanfang Med Univ, Nanfang Hosp, Dept Orthoped & Traumatol, Guangzhou 510515, Guangdong, Peoples R China
关键词
3D; acetabulum; bone screw; guide apparatu; HETEROTOPIC OSSIFICATION; OPERATIVE TREATMENT; SCREW FIXATION; LAG SCREW; SURGERY; GUIDANCE;
D O I
10.1097/MD.0000000000012134
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to analyze the feasibility and accuracy of a newly developed guide apparatus for the percutaneous retrograde lag screw fixation of posterior column of acetabular fractures. 3D pelvic models were reconstructed from the helical computed tomographic data of 33 adult patients using the Mimics 10.01 software. The virtual cylindrical implants were placed along a line passing through the central point of the ischial tuberosity and the midpoint between the most prominent point of anterior superior iliac spine and that of posterior superior iliac spine. Some anatomical parameters were then measured, based on which a guide apparatus was developed, and its safety and accuracy were experimentally validated with pelvic and cadaveric specimens. The screws were successfully placed in all of the 66 hemipelves. There was a significant difference between the male and female groups in the AB distance (156.26 +/- 7.28 mm and 151.38 +/- 8.11 mm), OI distance (139.53 +/- 7.56 mm and 125.15 +/- 11.17 mm), and diameter (12.19 +/- 1.97 mm and 10.19 +/- 2.14 mm) of the virtual cylindrical implants. This guide apparatus was proved effective for percutaneous retrograde lag screw fixation of posterior column acetabular fractures by the experiments with the pelvic and cadaveric specimens. Screw fixation of posterior column fractures via OI is safe and feasible. We designed a new percutaneous retrograde screw fixation guide apparatus to assist internal fixation of posterior column acetabular fracture.
引用
收藏
页数:7
相关论文
共 25 条
  • [21] The Kocher-Langenbeck approach combined with robot-aided percutaneous anterior column screw fixation for transverse-oriented acetabular fractures: a retrospective study
    Liu, Zhao-jie
    Gu, Ya
    Jia, Jian
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [22] The efficacy of posterior plating and anterior column screw fixation in the management of T-shaped acetabular fractures - CART analysis of prospective cohort study
    Hammad, A. S.
    El-Khadrawe, T. A.
    Waly, A. H.
    Abu-Sheasha, G. A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (03): : 680 - 686
  • [23] Is an Anatomical Suprapectineal Quadrilateral Surface Plate Superior to Previous Fixation Methods for Anterior Column-Posterior Hemitransverse Acetabular Fractures Typical in the Elderly?: A Biomechanical Study
    Kwak, Dae-Kyung
    Jang, Ji-Eun
    Kim, Won-Hyeon
    Lee, Sung -Jae
    Lee, Yongmin
    Yoo, Je-Hyun
    CLINICS IN ORTHOPEDIC SURGERY, 2023, 15 (02) : 182 - 191
  • [24] The Kocher-Langenbeck approach combined with robot-aided percutaneous anterior column screw fixation for transverse-oriented acetabular fractures: a retrospective study
    Zhao-jie Liu
    Ya Gu
    Jian Jia
    BMC Musculoskeletal Disorders, 23
  • [25] Design and application of a novel patient-specific 3D printed drill navigational guiding template in percutaneous thoracolumbar pedicle screw fixation: A cadaveric study
    Li, Zhonghai
    Xu, Dingjun
    Li, FengNing
    Liu, Mozhen
    Xu, Gang
    Yang, Ming
    JOURNAL OF CLINICAL NEUROSCIENCE, 2020, 73 : 294 - 298