Measurement of Core Body Temperature Using Graphene-Inked Infrared Thermopile Sensor

被引:44
作者
Chaglla, Jorge S. E. [1 ]
Celik, Numan [2 ]
Balachandran, Wamadeva [1 ]
机构
[1] Brunel Univ London, Dept Elect & Comp Engn, Kingston Lane, Uxbridge UB8 3PH, Middx, England
[2] Kings Coll London, Inst Pharmaceut Sci, London SE1 9NH, England
关键词
core body temperature; graphene ink; infrared thermopile sensor; distant measurement; ear canal sensing;
D O I
10.3390/s18103315
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Continuous and reliable measurements of core body temperature (CBT) are vital for studies on human thermoregulation. Because tympanic membrane directly reflects the temperature of the carotid artery, it is an accurate and non-invasive method to record CBT. However, commercial tympanic thermometers lack portability and continuous measurements. In this study, graphene inks were utilized to increase the accuracy of the temperature measurements from the ear by coating graphene platelets on the lens of an infrared thermopile sensor. The proposed ear-based device was designed by investigating ear canal geometry and developed with 3D printing technology using the Computer-Aided Design (CAD) Software, SolidWorks 2016. It employs an Arduino Pro Mini and a Bluetooth module. The proposed system runs with a 3.7 V, 850 mAh rechargeable lithium-polymer battery that allows long-term, continuous monitoring. Raw data are continuously and wirelessly plotted on a mobile phone app. The test was performed on 10 subjects under resting and exercising in a total period of 25 min. Achieved results were compared with the commercially available Braun Thermoscan, Original Thermopile, and Cosinuss One ear thermometers. It is also comprehended that such system will be useful in personalized medicine as wearable in-ear device with wireless connectivity.
引用
收藏
页数:8
相关论文
共 19 条
[1]  
Boano C.A., 2013, P 2013 IEEE INT C BO, P1
[2]  
Bock M, 2005, CAN J ANAESTH, V52, P1083, DOI 10.1007/BF03021609
[3]  
British Standards Institution, 2011, 6026872011 BRIT STAN
[4]  
Celik N., 2017, P IEEE SENS C SENS 2
[5]   Graphene-Enabled Electrodes for Electrocardiogram Monitoring [J].
Celik, Numan ;
Manivannan, Nadarajah ;
Strudwick, Andrew ;
Balachandran, Wamadeva .
NANOMATERIALS, 2016, 6 (09)
[6]   Graphene-based biosensors: methods, analysis and future perspectives [J].
Celik, Numan ;
Balachandran, Wamadeva ;
Manivannan, Nadarajah .
IET CIRCUITS DEVICES & SYSTEMS, 2015, 9 (06) :434-445
[7]   Science and technology roadmap for graphene, related two-dimensional crystals, and hybrid systems [J].
Ferrari, Andrea C. ;
Bonaccorso, Francesco ;
Fal'ko, Vladimir ;
Novoselov, Konstantin S. ;
Roche, Stephan ;
Boggild, Peter ;
Borini, Stefano ;
Koppens, Frank H. L. ;
Palermo, Vincenzo ;
Pugno, Nicola ;
Garrido, Jose A. ;
Sordan, Roman ;
Bianco, Alberto ;
Ballerini, Laura ;
Prato, Maurizio ;
Lidorikis, Elefterios ;
Kivioja, Jani ;
Marinelli, Claudio ;
Ryhaenen, Tapani ;
Morpurgo, Alberto ;
Coleman, Jonathan N. ;
Nicolosi, Valeria ;
Colombo, Luigi ;
Fert, Albert ;
Garcia-Hernandez, Mar ;
Bachtold, Adrian ;
Schneider, Gregory F. ;
Guinea, Francisco ;
Dekker, Cees ;
Barbone, Matteo ;
Sun, Zhipei ;
Galiotis, Costas ;
Grigorenko, Alexander N. ;
Konstantatos, Gerasimos ;
Kis, Andras ;
Katsnelson, Mikhail ;
Vandersypen, Lieven ;
Loiseau, Annick ;
Morandi, Vittorio ;
Neumaier, Daniel ;
Treossi, Emanuele ;
Pellegrini, Vittorio ;
Polini, Marco ;
Tredicucci, Alessandro ;
Williams, Gareth M. ;
Hong, Byung Hee ;
Ahn, Jong-Hyun ;
Kim, Jong Min ;
Zirath, Herbert ;
van Wees, Bart J. .
NANOSCALE, 2015, 7 (11) :4598-4810
[8]   Temperature regulation during exercise [J].
Gleeson, M .
INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1998, 19 :S96-S99
[9]   The usefulness of an earphone-type infrared tympanic thermometer for Intraoperative core temperature monitoring [J].
Kiya, Tomohiro ;
Yamakage, Michiaki ;
Hayase, Tomo ;
Satoh, Jun-Ichi ;
Namiki, Akiyoshi .
ANESTHESIA AND ANALGESIA, 2007, 105 (06) :1688-1692
[10]  
Lee W., 2016, P HUM FACT ERG SOC A, V2003, P1052, DOI DOI 10.1177/1541931213601244