Nuclear Probes and Intraoperative Gamma Cameras

被引:76
作者
Heller, Sherman [2 ]
Zanzonico, Pat [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[2] Montefiore Med Ctr, Dept Nucl Med, Bronx, NY 10467 USA
关键词
BREAST-CANCER SURGERY; SENTINEL LYMPH-NODES; HAND-HELD; LOCALIZATION; PERFORMANCE; LYMPHOSCINTIGRAPHY; POSITRON; BIOPSY; TUMORS; CDTE;
D O I
10.1053/j.semnuclmed.2010.12.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gamma probes are now an important, well-established technology in the management of cancer, particularly in the detection of sentinel lymph nodes. Intraoperative sentinel lymph node as well as tumor detection may be improved under some circumstances by the use of beta (negatron or positron), rather than gamma detection, because the very short range (similar to 1 mm or less) of such particulate radiations eliminates the contribution of confounding counts from activity other than in the immediate vicinity of the detector. This has led to the development of intraoperative beta probes. Gamma camera imaging also benefits from short source-to-detector distances and minimal overlying tissue, and intraoperative small field-of-view gamma cameras have therefore been developed as well. Radiation detectors for intraoperative probes can generally be characterized as either scintillation or ionization detectors. Scintillators used in scintillation-detector probes include thallium-doped sodium iodide, thallium- and sodium-doped cesium iodide, and cerium-doped lutecium orthooxy-silicate. Alternatives to inorganic scintillators are plastic scintillators, solutions of organic scintillation compounds dissolved in an organic solvent that is subsequently polymerized to form a solid. Their combined high counting efficiency for beta particles and low counting efficiency for 511-keV annihilation gamma-rays make plastic scintillators well-suited as intraoperative beta probes in general and positron probes in particular Semiconductors used in ionization-detector probes include cadmium telluride, cadmium zinc telluride, and mercuric iodide. Clinical studies directly comparing scintillation and semiconductor intraoperative probes have not provided a clear choice between scintillation and ionization detector-based probes. The earliest small field-of-view intraoperative gamma camera systems were hand-held devices having fields of view of only 1.5-2.5 cm in diameter that used conventional thallium-doped sodium iodide or sodium-doped cesium iodide scintillation detectors. Later units used 2-dimensional arrays (mosaics) of scintillation crystals connected to a position-sensitive photomultiplier tube and, more recently, semiconductors such as cadmium telluride or cadmium zinc telluride. The main problems with the early units were their very small fields of view and the resulting large number of images required to interrogate the surgical field and the difficulty in holding the device sufficiently still for the duration (up to 1 min) of the image acquisition. More recently, larger field-of-view (up to 5 x 5 cm) devices have developed which are attached to an articulating arm for easy and stable positioning. These systems are nonetheless fully portable and small enough overall to be accommodated in typical surgical suites. Semin Nucl Med 41:166-181 (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:166 / 181
页数:16
相关论文
共 61 条
[1]   Update on detection of sentinel lymph nodes in patients with breast cancer [J].
Aarsvold, JN ;
Alazraki, NP .
SEMINARS IN NUCLEAR MEDICINE, 2005, 35 (02) :116-128
[2]   Performance evaluation of a hand-held, semiconductor (CdZnTe)-based gamma camera [J].
Abe, A ;
Takahashi, N ;
Lee, J ;
Oka, T ;
Shizukuishi, K ;
Kikuchi, T ;
Inoue, T ;
Jimbo, M ;
Ryuo, H ;
Bickel, C .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (06) :805-811
[3]   Lymphoscintigraphy, the sentinel node concept, and the intraoperative gamma probe in melanoma, breast cancer, and other potential cancers [J].
Alazraki, NP ;
Eshima, D ;
Eshima, LA ;
Herda, SC ;
Murray, DR ;
Vansant, JP ;
Taylor, AT .
SEMINARS IN NUCLEAR MEDICINE, 1997, 27 (01) :55-67
[4]  
ANDERSON D.W., 1984, ABSORPTION IONIZING
[5]  
Attix F H, 1986, INTRO RADIOLOGICAL P
[6]   COMPARISON OF INVIVO SCINTILLATION PROBES AND GAMMA CAMERAS FOR DETECTION OF SMALL, DEEP TUMORS [J].
BARBER, HB ;
BARRETT, HH ;
WOOLFENDEN, JM ;
MYERS, KJ ;
HICKERNELL, TS .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) :727-739
[7]   COMPARISON OF NAL(TL), CDTE, AND HGL2 SURGICAL PROBES - PHYSICAL CHARACTERIZATION [J].
BARBER, HB ;
BARRETT, HH ;
HICKERNELL, TS ;
KWO, DP ;
WOOLFENDEN, JM ;
ENTINE, G ;
BACCASH, CO .
MEDICAL PHYSICS, 1991, 18 (03) :373-381
[8]   A method to evaluate intra-operative gamma probes for sentinel lymph node localisation [J].
Britten, AJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (02) :76-83
[9]  
Cherry S.R., 2003, Physics in Nuclear Medicine, V3rd, P325
[10]  
Chiesa C, 2001, EUR J NUCL MED, V28, P1110