Tracheal gas insufflation pressure control versus volume control ventilation - A lung model study

被引:20
作者
Imanaka, H
Kacmarek, RM
Ritz, R
Hess, D
机构
[1] MASSACHUSETTS GEN HOSP,RESP CARE DEPT LAB,BOSTON,MA 02114
[2] MASSACHUSETTS GEN HOSP,DEPT ANESTHESIA,BOSTON,MA 02114
[3] HARVARD UNIV,SCH MED,BOSTON,MA
关键词
D O I
10.1164/ajrccm.153.3.8630540
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Tracheal gas insufflation (TCI) has been recommended as an adjunct to mechanical ventilation in the presence of elevated Pa-CO2. Based on our initial clinical experience with continuous flow TGI and pressure control ventilation (PCV), we were concerned about elevation in peak airway pressure as TCI was applied. In a lung model, we evaluated the effects of continuous flow TCI during both PCV and volume control ventilation (VCV). A single compartment lung model was configured with an artificial trachea into which an 8-mm endotracheal tube was positioned. TCI was established with a 16-G catheter positioned 2 cm beyond the tip of the endotracheal tube. Ventilation was provided by a Puritan-Bennett 7200ae ventilator with PCV 20 cm H2O or VCV with a tidal volume (VT) similar to that with PCV. A rate of 15 breaths/min and PEEP of 10 cm H2O were used throughout. Inspiratory times (TI) of 1.0, 1.5, 2.0, and 2.5 s were used with TCI of 0, 4, 8, and 12 L/min. Lung model compliance (ml/cm H2O) and resistance (cm H(2)0/L/s) combinations of 20/20, 20/5, and 50/20 were used. Auto-PEEP, VT, and peak alveolar and airway opening pressures increased as TCI and TI increased, regardless of lung mechanics settings (p < 0.01). All increases were greater with VCV than PCV (p < 0.05). Continuous flow TCI with both PCV and VT-uncorrected VCV may result in marked increases in VT and system pressures, especially at long TI.
引用
收藏
页码:1019 / 1024
页数:6
相关论文
共 19 条
[1]   MODES OF TRACHEAL GAS INSUFFLATION - COMPARISON OF CONTINUOUS AND PHASE-SPECIFIC GAS INJECTION IN NORMAL DOGS [J].
BURKE, WC ;
NAHUM, A ;
RAVENSCRAFT, SA ;
NAKOS, G ;
ADAMS, AB ;
MARCY, TW ;
MARINI, JJ .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 148 (03) :562-568
[2]   ADVERSE-EFFECTS OF LARGE TIDAL VOLUME AND LOW PEEP IN CANINE ACID ASPIRATION [J].
CORBRIDGE, TC ;
WOOD, LDH ;
CRAWFORD, GP ;
CHUDOBA, MJ ;
YANOS, J ;
SZNAJDER, JI .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 142 (02) :311-315
[3]   ROLE OF TIDAL VOLUME, FRC, AND END-INSPIRATORY VOLUME IN THE DEVELOPMENT OF PULMONARY-EDEMA FOLLOWING MECHANICAL VENTILATION [J].
DREYFUSS, D ;
SAUMON, G .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 148 (05) :1194-1203
[4]   LOW MORTALITY-RATE IN ADULT-RESPIRATORY-DISTRESS-SYNDROME USING LOW-VOLUME, PRESSURE-LIMITED VENTILATION WITH PERMISSIVE HYPERCAPNIA - A PROSPECTIVE-STUDY [J].
HICKLING, KG ;
WALSH, J ;
HENDERSON, S ;
JACKSON, R .
CRITICAL CARE MEDICINE, 1994, 22 (10) :1568-1578
[5]   LOW MORTALITY ASSOCIATED WITH LOW-VOLUME PRESSURE LIMITED VENTILATION WITH PERMISSIVE HYPERCAPNIA IN SEVERE ADULT RESPIRATORY-DISTRESS SYNDROME [J].
HICKLING, KG ;
HENDERSON, SJ ;
JACKSON, R .
INTENSIVE CARE MEDICINE, 1990, 16 (06) :372-377
[6]  
JONSON B, 1990, EUR RESPIR J, V3, P1202
[7]   CONSTANT-FLOW VENTILATION OF APNEIC DOGS [J].
LEHNERT, BE ;
OBERDORSTER, G ;
SLUTSKY, AS .
JOURNAL OF APPLIED PHYSIOLOGY, 1982, 53 (02) :483-489
[8]   A GENERAL MATHEMATICAL-MODEL FOR RESPIRATORY DYNAMICS RELEVANT TO THE CLINICAL SETTING [J].
MARINI, JJ ;
CROOKE, PS .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 147 (01) :14-24
[9]  
Marini JJ, 1994, PRINCIPLES PRACTICE, P305
[10]   LUNG-MECHANICS AND GAS-EXCHANGE DURING PRESSURE-CONTROL VENTILATION IN DOGS - AUGMENTATION OF CO2 ELIMINATION BY AN INTRATRACHEAL CATHETER [J].
NAHUM, A ;
BURKE, WC ;
RAVENSCRAFT, SA ;
MARCY, TW ;
ADAMS, AB ;
CROOKE, PS ;
MARINI, JJ .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1992, 146 (04) :965-973