Measurement of acetylcholine-induced endothelium-derived nitric oxide in aorta using a newly developed catheter-type nitric oxide sensor

被引:27
作者
Mochizuki, S
Miyasaka, T
Goto, M
Ogasawara, Y
Yada, T
Akiyama, M
Neishi, Y
Toyoda, T
Tomita, J
Koyama, Y
Tsujioka, K
Kajiya, F
Akasaka, T
Yoshida, K
机构
[1] Kawasaki Med Sch, Dept Med Engn, Kurashiki, Okayama 7010192, Japan
[2] Kawasaki Med Sch, Dept Physiol, Okayama, Japan
[3] Kawasaki Med Sch, Div Cardiovasc Med, Okayama, Japan
[4] Okayama Univ, Grad Sch Med & Dent, Dept Cardiovasc Physiol, Okayama, Japan
关键词
aorta; catheter; endothelium-derived nitric oxide; sensor;
D O I
10.1016/S0006-291X(03)00985-9
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Intra-aortic measurement of nitric oxide (NO) would provide valuable insights into NO bioavailability in systemic circulation and vascular endothelial function. In the present study, we thus developed a catheter-type NO sensor to measure intra-aortic NO concentration in vivo. An NO sensor was encased and fixed in a 4-Fr catheter. The sensor was then located in the thoracic aorta via the femoral artery through a 7-Fr catheter to measure intra-aortic plasma NO concentration in vivo in anesthetized dogs. Infusion of acetylcholine (10 mug/kg) increased base-to-peak plasma NO level in the aorta by 2.4 +/- 0.4 nM (n = 7). After 20-min infusion of N-G-methyl-L-arginine (NO synthase inhibitor), changes in plasma NO concentration in response to acetylcholine were attenuated significantly (1.8 +/- 10.4 nM, P < 0.003, n = 7). In conclusion, the newly developed catheter-type NO sensor successfully measured acetylcholine-induced changes in intra-aortic plasma concentration of endothelium-derived NO in vivo and demonstrated applicability to direct evaluation of intravascular NO bioavailability. (C) 2003 Elsevier Science (USA). All rights reserved.
引用
收藏
页码:505 / 508
页数:4
相关论文
共 13 条
  • [1] Endothelial function: From vascular biology to clinical applications
    Behrendt, D
    Ganz, P
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (10C) : 40L - 48L
  • [2] Endothelial dysfunction in cardiovascular diseases - The role of oxidant stress
    Cai, H
    Harrison, DG
    [J]. CIRCULATION RESEARCH, 2000, 87 (10) : 840 - 844
  • [3] Gerova M, 1998, PHYSIOL RES, V47, P169
  • [4] KELM M, 1996, METHODS NITRIC OXIDE, P47
  • [5] NITRIC-OXIDE RELEASE FROM A SINGLE CELL MEASURED INSITU BY A PORPHYRINIC-BASED MICROSENSOR
    MALINSKI, T
    TAHA, Z
    [J]. NATURE, 1992, 358 (6388) : 676 - 678
  • [6] Evaluation of basic performance and applicability of a newly developed in vivo nitric oxide sensor
    Mochizuki, S
    Himi, N
    Miyasaka, T
    Nakamoto, H
    Takemoto, M
    Hirano, K
    Tsujioka, K
    Ogasawara, Y
    Kajiya, F
    [J]. PHYSIOLOGICAL MEASUREMENT, 2002, 23 (02) : 261 - 268
  • [7] Prognostic significance of endothelial dysfunction in hypertensive patients
    Perticone, F
    Ceravolo, R
    Pujia, A
    Ventura, G
    Iacopino, S
    Scozzafava, A
    Ferraro, A
    Chello, M
    Mastroroberto, P
    Verdecchia, P
    Schillaci, G
    [J]. CIRCULATION, 2001, 104 (02) : 191 - 196
  • [8] Mechanical transduction of nitric oxide synthesis in the beating heart
    Pinsky, DJ
    Patton, S
    Mesaros, S
    Brovkovych, V
    Kubaszewski, E
    Grunfeld, S
    Malinski, T
    [J]. CIRCULATION RESEARCH, 1997, 81 (03) : 372 - 379
  • [9] Evidence for in vivo transport of bioactive nitric oxide in human plasma
    Rassaf, T
    Preik, M
    Kleinbongard, P
    Lauer, T
    Heiss, C
    Strauer, BE
    Feelisch, M
    Kelm, M
    [J]. JOURNAL OF CLINICAL INVESTIGATION, 2002, 109 (09) : 1241 - 1248
  • [10] Plasma nitrosothiols contribute to the systemic vasodilator effects of intravenously applied NO - Experimental and clinical study on the fate of NO in human blood
    Rassaf, T
    Kleinbongard, P
    Preik, M
    Dejam, A
    Gharini, P
    Lauer, T
    Erckenbrecht, J
    Duschin, A
    Schulz, R
    Heusch, G
    Feelisch, M
    Kelm, M
    [J]. CIRCULATION RESEARCH, 2002, 91 (06) : 470 - 477