Comparison BIPM.RI(I)-K8 of high dose-rate Ir-192 brachytherapy standards for reference air kerma rate of the VSL and the BIPM

被引:0
作者
Alvarez, J. T. [1 ,4 ]
de Pooter, J. A. [2 ]
Andersen, C. [3 ]
Aalbers, A. H. L. [2 ]
Allisy-Roberts, P. J. [4 ]
Kessler, C. [4 ]
机构
[1] ININ, Ocoyoacac, Mexico
[2] Dutch Metrol Inst, VSL, Delft, Netherlands
[3] Riso High Dose Reference Lab, Roskilde, Denmark
[4] Bur Int Poids & Mesures, F-92312 Sevres, France
关键词
Radiotherapy;
D O I
10.1088/0026-1394/51/1A/06022
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An indirect comparison of the standards for reference air kerma rate for Ir-192 high dose rate (HDR) brachytherapy sources of the Dutch Metrology Institute (VSL), The Netherlands, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the VSL in November 2009. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the VSL and the BIPM standards for reference air kerma rate, is 0.9873 with a combined standard uncertainty of 0.0061.
引用
收藏
页数:10
相关论文
共 11 条
[1]  
Allisy-Roberts P.J., 2007, METROLOGIA TECH S, V44, P06006
[2]  
Allisy-Roberts P.J., 2009, BIPM0904
[3]  
[Anonymous], 2008, JCGM
[4]  
Burns D.T., 2009, KEY COMP BIPM RI I K
[5]  
KCDB, 2014, BIPM KEY COMP DAT BA
[6]   On the accuracy of techniques for obtaining the calibration coefficient NK of 192Ir HDR brachytherapy sources [J].
Mainegra-Hing, Ernesto ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2006, 33 (09) :3340-3347
[7]  
Petersen J.J., 1994, SEI9401 NMIVSL
[8]  
Petersen J.J., 1994, SEI9402 NMIVSL
[9]   Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty [J].
van Dijk, E ;
Kolkman-Deurloo, IKK ;
Damen, PMG .
MEDICAL PHYSICS, 2004, 31 (10) :2826-2833
[10]  
van Dijk E, 2007, VSLESLIO20073 NMI