Towards real-time thermometry using simultaneous multislice MRI

被引:15
作者
Borman, P. T. S. [1 ,2 ]
Bos, C. [2 ]
de Boorder, T. [3 ]
Raaymakers, B. W. [1 ]
Moonen, C. T. W. [2 ]
Crijns, S. P. M. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Imaging Div, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
[3] Univ Med Ctr Utrecht, Med Technol & Clin Phys, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
关键词
MR-guided therapy; thermal therapy; MR thermometry; multiband imaging; simultaneous multislice imaging; SIGNAL-TO-NOISE; THERMAL THERAPY; TEMPERATURE; HIFU; RECONSTRUCTION; ABLATION; BRAIN; SHIFT;
D O I
10.1088/0031-9155/61/17/N461
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
MR-guided thermal therapies, such as high-intensity focused ultrasound (MRgHIFU) and laser-induced thermal therapy (MRgLITT) are increasingly being applied in oncology and neurology. MRI is used for guidance since it can measure temperature noninvasively based on the proton resonance frequency shift (PRFS). For therapy guidance using PRFS thermometry, high temporal resolution and large spatial coverage are desirable. We propose to use the parallel imaging technique simultaneous multislice (SMS) in combination with controlled aliasing (CAIPIRINHA) to accelerate the acquisition. We compare this with the sensitivity encoding (SENSE) acceleration technique. Two experiments were performed to validate that SMS can be used to increase the spatial coverage or the temporal resolution. The first was performed in agar gel using LITT heating and a gradient-echo sequence with echo-planar imaging (EPI), and the second was performed in bovine muscle using HIFU heating and a gradient-echo sequence without EPI. In both experiments temperature curves from an unaccelerated scan and from SMS, SENSE, and SENSE/ SMS accelerated scans were compared. The precision was quantified by a standard deviation analysis of scans without heating. Both experiments showed a good agreement between the temperature curves obtained from the unaccelerated, and SMS accelerated scans, confirming that accuracy was maintained during SMS acceleration. The standard deviations of the temperature measurements obtained with SMS were significantly smaller than when SENSE was used, implying that SMS allows for higher acceleration. In the LITT and HIFU experiments SMS factors up to 4 and 3 were reached, respectively, with a loss of precision of less than a factor of 3. Based on these results we conclude that SMS acceleration of PRFS thermometry is a valuable addition to SENSE, because it allows for a higher temporal resolution or bigger spatial coverage, with a higher precision.
引用
收藏
页码:N461 / N477
页数:17
相关论文
共 37 条
  • [21] 2-S
  • [22] Quesson B, 2000, J MAGN RESON IMAGING, V12, P525, DOI 10.1002/1522-2586(200010)12:4<525::AID-JMRI3>3.0.CO
  • [23] 2-V
  • [24] Thermal ablative treatment of uterine fibroids
    Quinn, Stephen Derek
    Gedroyc, Wladyslaw M.
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2015, 31 (03) : 272 - 279
  • [25] MR Thermometry in the Human Prostate Gland at 3.0T for Transurethral Ultrasound Therapy
    Ramsay, Elizabeth
    Mougenot, Charles
    Kohler, Max
    Bronskill, Michael
    Klotz, Laurence
    Haider, Masoom A.
    Chopra, Rajiv
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (06) : 1564 - 1571
  • [26] Temperature measurement artefacts of thermocouples and fluoroptic probes during laser irradiation at 810 nm
    Reid, AD
    Gertner, MR
    Sherar, MD
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (06) : N149 - N157
  • [27] MR thermometry
    Rieke, Viola
    Pauly, Kim Butts
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 376 - 390
  • [28] Robust Adaptive Extended Kalman Filtering for Real Time MR-Thermometry Guided HIFU Interventions
    Roujol, Sebastien
    de Senneville, Baudouin Denis
    Hey, Silke
    Moonen, Chrit
    Ries, Mario
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) : 533 - 542
  • [29] Time Efficient Design of Multi dimensional RF Pulses: Application of a Multi shift CGLS Algorithm
    Sbrizzi, Alessandro
    Hoogduin, Hans
    Lagendijk, Jan J.
    Luijten, Peter
    Sleijpen, Gerard L. G.
    van den Berg, Cornelis A. T.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (03) : 879 - 885
  • [30] Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty
    Setsompop, Kawin
    Gagoski, Borjan A.
    Polimeni, Jonathan R.
    Witzel, Thomas
    Wedeen, Van J.
    Wald, Lawrence L.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) : 1210 - 1224