Lumen claims of the STERRAD 100NX sterilizer: Testing performance limits when processing equipment containing long, narrow lumens

被引:6
作者
Diab-Elschahawi, Magda [1 ]
Blacky, Alexander [1 ]
Bachhofner, Nicole [1 ]
Koller, Walter [1 ]
机构
[1] Med Univ Vienna, Clin Inst Hosp Hyg, Vienna Gen Hosp, A-1090 Vienna, Austria
关键词
STERRAD; 100NX; flex scope cycle; low temperature H(2)O(2) gas plasma sterilization; GAS PLASMA STERILIZATION; ENDOSCOPE DECONTAMINATION; INFECTION;
D O I
10.1016/j.ajic.2011.01.010
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: According to manufacturers information, the STERRAD 100NX sterilizer-a low temperature H(2)O(2) gas plasma sterilizer-can adequately process single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a maximum length of 500 mm using standard cycle sterilizing conditions. The aim of this study was to qualify the performance of this H(2)O(2) gas plasma sterilizer under different experimental settings representing worst case conditions. Methods: Inoculated carriers were placed at the midpoint position of specified lumens and then submitted to flex scope sterilizing conditions. To simulate insufficient cleaning or crystalline residues, we added organic and inorganic challenges to our inoculated carriers. Results: For experiments done with unchallenged carriers, quantitative analysis reached a log(10) reduction rate of >= 5.71, whereas qualitative results showed no growth in 24 out of 30 biologic indicators tested using flex scope half cycle conditions. Any additional kind of challenge significantly impaired the sterilization outcome. Conclusion: The findings of our current study emphasize the importance of a thorough validated cleaning of medical devices as well as timing for cleaning and decontamination before being exposed to the H(2)O(2) sterilization process and, furthermore, the need for strict adherence to manufacturer's recommendations.
引用
收藏
页码:770 / 774
页数:5
相关论文
共 22 条
  • [1] Cleaning efficacy of medical device washers in North American healthcare facilities
    Alfa, M. J.
    Olson, N.
    Al-Fadhaly, A.
    [J]. JOURNAL OF HOSPITAL INFECTION, 2010, 74 (02) : 168 - 177
  • [2] *AOAC, 2000, OFF METH AN AOAC, pCH6
  • [3] *ASP, 2008, STERRAD 100NX US GUI
  • [4] ENDOSCOPE DECONTAMINATION - WHERE DO WE GO FROM HERE
    BABB, JR
    BRADLEY, CR
    [J]. JOURNAL OF HOSPITAL INFECTION, 1995, 30 : 543 - 551
  • [5] BERUBE R, 2001, DISINFECTION STERILI, P1367
  • [6] ENDOSCOPE DECONTAMINATION - AUTOMATED VS MANUAL
    BRADLEY, CR
    BABB, JR
    [J]. JOURNAL OF HOSPITAL INFECTION, 1995, 30 : 537 - 542
  • [7] Natural bioburden levels detected on rigid lumened medical devices before and after cleaning
    Chan-Meyers, H
    McAlister, D
    Antonoplos, P
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 1997, 25 (06) : 471 - 476
  • [8] Chu N S, 2000, Gastrointest Endosc Clin N Am, V10, P233
  • [9] Re-evaluation of current A0 value recommendations for thermal disinfection of reusable human waste containers based on new experimental data
    Diab-Elschahawi, M.
    Fuernkranz, U.
    Blacky, A.
    Bachhofner, N.
    Koller, W.
    [J]. JOURNAL OF HOSPITAL INFECTION, 2010, 75 (01) : 62 - 65
  • [10] *DTSCH GES KRANK, 1995, HYG MED, V20, P52