Mistakes encountered during automatic peak identification in low beam energy X-ray Microanalysis

被引:30
作者
Newbury, Dale E. [1 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
关键词
energy dispersive X-ray spectrometry; peak identification; peak fitting; qualitative analysis; scanning electron microscopy; X-ray microanalysis;
D O I
10.1002/sca.20009
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Automated peak identification in electron beam excited X-ray microanalysis with energy dispersive X-ray spectrometry (EDS) is subject to occasional mistakes even on well-separated, high-intensity peaks arising from major constituents. The problem is exacerbated when analysis conditions are restricted to operation in the "low beam energy scanning electron microscopy" (i.e. "low voltage scanning electron microscopy" or LVSEM) regime where the incident beam energy is 5 keV or less. These low beam energy microanalysis conditions force the analyst to use low fluorescence yield L-shell and M-shell peaks rather than higher yield K-shell and L-shell peaks typically selected for elements of intermediate and high atomic number under conventional high beam energy ( > 10 keV) conditions. Misidentifications can arise in automated peak identification procedures when only a single energy channel is used to characterize an EDS peak. The effect of the EDS measurement process is to convolve the closely spaced L alpha-L beta and M alpha-M beta peaks into a single peak with a peak channel shift of 20 eV or more from the L alpha or M alpha value, which is typically sought in an X-ray database. An extensive list of problem situations encountered in low beam energy microanalysis is presented based upon observed peak identification mistakes as well as likely troublesome situations based upon proximity in peak energy. Robust automatic peak identification requires implementation of peak fitting that utilizes the full peak shape.
引用
收藏
页码:137 / 151
页数:15
相关论文
共 8 条
[1]  
Bevington P., 2002, Data Reduction and Error Analysis for the Physical Sciences, V3rd ed.
[2]  
FIORI C, 1991, DESKTOP SPECTRUM ANA
[3]  
Goldstein J. I., 2003, SCANNING ELECT MICRO, P355
[4]  
GOLDSTEIN JI, 1992, SCANNING ELECTRON MI, P377
[5]   Misidentification of major constituents by automatic qualitative energy dispersive X-ray microanalysis: A problem that threatens the credibility of the analytical community [J].
Newbury, DE .
MICROSCOPY AND MICROANALYSIS, 2005, 11 (06) :545-561
[6]  
Schamber F. H., 1981, Energy Dispersive X-Ray Spectrometry. Proceedings of a Workshop (NBS-SP-604), P193
[7]   EMPIRICAL METHOD FOR ELECTRON MICROANALYSIS [J].
ZIEBOLD, TO ;
OGILVIE, RE .
ANALYTICAL CHEMISTRY, 1964, 36 (02) :322-&
[8]   PRECISION AND SENSITIVITY IN ELECTRON MICROPROBE ANALYSIS [J].
ZIEBOLD, TO .
ANALYTICAL CHEMISTRY, 1967, 39 (08) :858-&