The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs

被引:32
作者
Bert, JM [1 ]
Reuben, J [1 ]
Kelly, F [1 ]
Gross, M [1 ]
Elting, J [1 ]
机构
[1] Univ Minnesota, Sch Med, St Paul, MN 55108 USA
关键词
modular tibial insert; polyethylene failure; early exchange;
D O I
10.1016/S0883-5403(98)80001-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
One of the primary reasons for utilizing modular tibial polyethylene inserts (MTPI) at the time of total knee arthroplasty is to have the ability to simply exchange the polyethylene at the time of revision surgery when polyethylene failure has occurred. During a 2-year period from January 1993 to December 1994, 62 revision total knee arthroplasties were reviewed from five different institutions in North America, which were performed secondary to modular tibial insert failure. In 55 cases (88.7%), significant scoring and/or damage to the femoral and/or tibial components occurred necessitating revision of one or both components. This series does not support the premise that polyethylene exchange is common at the time of revision surgery for MTPI failure. Of the patients 88.7% had MTPI failure resulted in femoral and/or tibial component revision secondary to surface damage to the femoral component and/or tibial component baseplate. In order to avoid femoral and tibial component wear secondary to MTPI failure, early recognition is necessary to allow the much simpler operation consisting of polyethylene liner exchange only.
引用
收藏
页码:609 / 614
页数:6
相关论文
共 15 条
  • [1] BARTLEY RE, 1994, CLIN ORTHOP RELAT R, P18
  • [2] BLOEBAUM, 1991, CLIN ORTHOP, V269, P128
  • [3] POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED
    BLUNN, GW
    JOSHI, AB
    LILLEY, PA
    ENGELBRECHT, E
    RYD, L
    LIDGREN, L
    HARDINGE, K
    NIEDER, E
    WALKER, PS
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03): : 247 - 255
  • [4] CHENG DC, 1995, BIO MAT, V16, P219
  • [5] CHILLAG KJ, 1991, CLIN ORTHOP RELAT R, P261
  • [6] COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
  • [7] POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES
    ENGH, GA
    DWYER, KA
    HANES, CK
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 9 - 17
  • [8] FENG EL, 1994, CLIN ORTHOP RELAT R, P60
  • [9] HECK DA, 1992, ORTHOPEDICS, V15, P23
  • [10] POLYETHYLENE WEAR IN UNCEMENTED KNEE REPLACEMENTS
    JONES, SMG
    PINDER, IM
    MORAN, CG
    MALCOLM, AJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 18 - 22