AC-DC Calibrations With a Pulse-driven AC Josephson Voltage Standard Operated in a Small Cryostat

被引:0
作者
van den Brom, Helko E. [1 ]
Kieler, Oliver F. O. [2 ]
Bauer, Stephan [2 ]
Houtzager, Ernest [1 ]
机构
[1] VSL Dutch Metrol Inst, Thijsseweg 11, NL-2629 JA Delft, Netherlands
[2] Phys Tech Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
来源
2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016) | 2016年
关键词
Measurement standards; precision measurements; AC voltage measurement; pulse-driven AC Josephson voltage standard;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
AC-DC measurements have been performed using a pulse-driven AC Josephson Voltage Standard with a small cryostat. The deviation at higher frequency was an order of magnitude smaller than was observed before with a normal cryostat. Furthermore, the influence of other cables and equipment was investigated. The results support the explanation of the frequency dependence in terms of standing waves.
引用
收藏
页数:2
相关论文
共 5 条
  • [1] International comparison of quantum AC voltage standards for frequencies up to 100 kHz
    Filipski, Piotr S.
    van den Brom, Helko E.
    Houtzager, Ernest
    [J]. MEASUREMENT, 2012, 45 (09) : 2218 - 2225
  • [2] Experimental Determination of the Voltage Lead Error in an AC Josephson Voltage Standard
    Filipski, Piotr S.
    Boecker, Michael
    Benz, Samuel P.
    Burroughs, Charles J., Jr.
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (07) : 2387 - 2392
  • [3] Kieler OF, 2010, 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, P157, DOI 10.1109/CPEM.2010.5544751
  • [4] Voltage lead corrections for a pulse-driven ac Josephson voltage standard
    van den Brom, Helko E.
    Houtzager, Ernest
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (12)
  • [5] van den Brom Helko E., 2016, CPEM C UNPUB