A simple method of measuring the wear of explanted acetabular component inserts

被引:0
作者
Krakow, L. [1 ]
Klockow, A. [1 ,2 ]
Roehner, E. [1 ]
Brodt, S. [1 ]
Eijer, H. [1 ,3 ]
Bossert, J. [1 ,4 ]
Matziolis, G. [1 ]
机构
[1] Univ Hosp Jena, Orthoped Dept, Campus Eisenberg, Jena, Germany
[2] Mathys AG, Dev Hip, Bettlach, Switzerland
[3] Spital Emmental, Dept Orthoped, Emmental, Switzerland
[4] Friedrich Schiller Univ Jena, Otto Schott Inst Mat Sci, Jena, Germany
来源
BONE & JOINT RESEARCH | 2017年 / 6卷 / 09期
关键词
Wear; Acetabular component; Total hip arthroplasty; TOTAL HIP-ARTHROPLASTY; VOLUMETRIC WEAR; POLYETHYLENE WEAR; LINEAR WEAR; IN-VIVO; CUPS; VALIDATION; CRACKING;
D O I
10.1302/2046-3758.69.BJR-2016-0249.R1
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Objectives The determination of the volumetric polyethylene wear on explanted material requires complicated equipment, which is not available in many research institutions. Our aim in this study was to present and validate a method that only requires a set of polyetheretherketone balls and a laboratory balance to determine wear. Methods The insert to be measured was placed on a balance, and a ball of the appropriate diameter was inserted. The cavity remaining between the ball and insert caused by wear was filled with contrast medium and the weight of the contrast medium was recorded. The volume was calculated from the known density of the liquid. The precision, inter-and intraobserver reliability, were determined by four investigators on four days using nine inserts with specified wear (0.094 ml to 1.626 ml), and the intra-class correlation coefficient was calculated. The feasibility of using this method in routine clinical practice and the time required for measurement were tested on 84 explanted inserts by one investigator. Results In order to get the mean for all investigators and determinations, the deviation between the measured and specified wear was -0.08 ml (sd 0.12; -0.21 to 0.11). The interobserver reliability was 0.989 ml (95% confidence interval (CI) 0.964 to 0.997) and the intraobserver reliability was 0.941 for observer 1 (95% CI 0.846 to 0.985), 0.983 for observer 2 (95% CI 0.956 to 0.995), 0.939 for observer 3 (95% CI 0.855 to 0.984), and 0.934 for observer 4 (95% CI 0.790 to 0.984). The mean time required to examine the samples was two minutes (sd 2; 1 to 5). Conclusion The method presented here was shown to be sufficiently precise for many settings and is a cost-effective and quick method of determining the volumetric wear of explanted acetabular components. However, the measurement of wear for scientific purposes will probably continue to involve more accurate and dedicated laboratory equipment.
引用
收藏
页码:530 / 534
页数:5
相关论文
共 28 条
  • [1] Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners
    Bowden, AE
    Kurtz, SM
    Edidin, AA
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 75B (01) : 205 - 209
  • [2] Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners
    Bradford, L
    Baker, DA
    Graham, J
    Chawan, A
    Ries, MD
    Pruitt, LA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) : 1271 - 1282
  • [3] Wear of retrieved cemented polyethylene acetabula with alumina femoral heads
    Bragdon, CR
    Jasty, M
    Kawate, K
    McGrory, BJ
    Elder, JR
    Lowenstein, J
    Harris, WH
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (02) : 119 - 125
  • [4] Brummitt K, 1996, Proc Inst Mech Eng H, V210, P187, DOI 10.1243/PIME_PROC_1996_210_411_02
  • [5] CHARNLEY J, 1975, CLIN ORTHOP RELAT R, P170
  • [6] Wear analysis of failed acetabular polyethylene - A Comparison of analytical methods
    Chuter, G. S. J.
    Cloke, D. J.
    Mahomed, A.
    Partington, P. F.
    Green, S. M.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (02): : 273 - 279
  • [7] Accuracy and Precision of Two Computer-Assisted Methods of Radiographic Wear Measurement in Total Hip Arthroplasty
    Crockarell, John R., Jr.
    Snearly, Christine M.
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (01) : 37 - 40
  • [8] Center index methodan alternative for wear measurements with radiostereometry (RSA)
    Dahl, Jon
    Figved, Wender
    Snorrason, Finnur
    Nordsletten, Lars
    Rohrl, Stephan M.
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (03) : 480 - 484
  • [9] DORR LD, 1995, CLIN ORTHOP RELAT R, P191
  • [10] The measurement of creep in ultrahigh molecular weight polyethylene - A comparison of conventional versus highly cross-linked polyethylene
    Estok, DM
    Bragdon, CR
    Plank, GR
    Huang, A
    Muratoglu, OK
    Harris, WH
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (02) : 239 - 243