Dynode-timing method for PET block detectors

被引:1
|
作者
Bauer, Florian [1 ,2 ]
Zhang, Nan [2 ]
Schmand, Matthias [2 ]
Loope, Michael [2 ]
Eriksson, Lars [1 ,2 ]
Aykac, Mehmet [2 ]
机构
[1] Univ Stockholm, Dept Phys, SE-10691 Stockholm, Sweden
[2] Siemens Med Sol, Mol Imaging, Knoxville, TN 37932 USA
关键词
detectors; dynode-timing; high resolution; positron emission tomography (PET); small animal; SPECT;
D O I
10.1109/TNS.2007.910879
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The focus of this paper is the investigation of a new dynode-timing technique optimized for PET block detectors. This method allows utilization of dynode signals from single but especially multiple photo-multiplier tubes (PMTs), operated with negative high-voltage. The technique will provide an event-timing trigger without deteriorating the anode signal. A printed circuit board has been developed and built for this investigation. Benchmark measurements have been performed, comparing timing of the anode signal with timing of the inverted last-dynode signal and timing of the dynode signal extracted via a newly developed LVPECL-logic based board. Timing measurements were performed with plastic as well as LSO scintillators. From single PMT measurements we find a 30 ps improvement with the dynode-timing method compared to the standard anode timing with two Photonis XP2020Q PMTs with LSO (10 mm x 10 mm x 10 ram). For a quad-PMT block detector, assembled of four Hamamatsu R9800 with a Hi-Rez block, the timing-resolution improves similar to 10%, by 43 ps compared to the standard anode timing.
引用
收藏
页码:451 / 456
页数:6
相关论文
共 50 条
  • [41] Timing Estimation and Limits in TOF-PET Detectors Producing Prompt Photons
    Loignon-Houle, Francis
    Toussaint, Maxime
    Bertrand, Emilie
    Lemyre, Felix Camirand
    Lecomte, Roger
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (07) : 692 - 703
  • [42] Analysis of timing error detectors for orthogonal space-time block codes
    Dmochowski, Pawel A.
    McLane, Peter J.
    2007 IEEE INTERNATIONAL CONFERENCE ON COMMUNICATIONS, VOLS 1-14, 2007, : 2949 - 2955
  • [43] A new method for depth of interaction determination in PET detectors
    Pizzichemi, M.
    Stringhini, G.
    Niknejad, T.
    Liu, Z.
    Lecoq, P.
    Tavernier, S.
    Varela, J.
    Paganoni, M.
    Auffray, E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (12): : 4679 - 4698
  • [44] TIMING IMPROVED BY THE USE OF DYNODE SIGNALS STUDIED WITH DIFFERENT SCINTILLATORS AND PHOTOMULTIPLIERS
    BENGTSON, B
    MOSZYNSKI, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 204 (01): : 129 - 140
  • [45] Investigation of analog charge multiplexing schemes for SiPM based PET block detectors
    Downie, Evan
    Yang, Xin
    Peng, Hao
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (11): : 3943 - 3964
  • [46] High speed position-decoding electronics for BGO block detectors in PET
    Li, Hongdi
    Wong, Wai-Hoi
    Uribe, Jorge
    Baghaei, Hossain
    Zhang, Nan
    Wang, Junming
    Wang, Yu.
    IEEE Transactions on Nuclear Science, 2000, 47 (3 III) : 1006 - 1010
  • [47] Bipolar analog signal multiplexing for position-sensitive PET block detectors
    Yoon, Hyun Suk
    Lee, Jae Sung
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (24): : 7835 - 7846
  • [48] CNN-based PET sinogram repair to mitigate defective block detectors
    Whiteley, William
    Gregor, Jens
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (23):
  • [49] A ROTATING PET SCANNER USING BGO BLOCK DETECTORS - DESIGN, PERFORMANCE AND APPLICATIONS
    TOWNSEND, DW
    WENSVEEN, M
    BYARS, LG
    GEISSBUHLER, A
    TOCHONDANGUY, HJ
    CHRISTIN, A
    DEFRISE, M
    BAILEY, DL
    GROOTOONK, S
    DONATH, A
    NUTT, R
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (08) : 1367 - 1376
  • [50] High speed position-decoding electronics for BGO block detectors in PET
    Li, HD
    Wong, WH
    Uribe, J
    Baghaei, H
    Zhang, N
    Wang, JM
    Wang, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 1006 - 1010