Scatter correction for clinical cone beam CT breast imaging based on breast phantom studies

被引:11
作者
Cai, Weixing [1 ]
Ning, Ruola [1 ]
Conover, David [1 ]
机构
[1] Univ Rochester, Med Ctr, Dept Imaging Sci, Rochester, NY 14642 USA
关键词
Cone beam CT; scatter correction; breast imaging; X-RAY SCATTER; AIR GAPS; ALGORITHM; RADIATION; REMOVAL;
D O I
10.3233/XST-2010-0280
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In flat-panel detector-based cone beam CT breast imaging (CBCTBI) systems, scattering is an important factor that degrades image quality. It is not practical to measure the scattering profiles of a breast for all view angles in a patient study, but it is possible to develop a method to estimate the scattering profiles based on information acquired from breast phantom studies. A new scattering correction method is proposed for clinical CBCTBI in this study. The scattering profiles of three anthropomorphic uncompressed breast phantoms of different sizes were thoroughly investigated, and the results indicated that though phantom size differed, the scattering profiles were mainly determined by local breast diameters, which are the approximate diameters of coronal slices that are perpendicular to the nipple-to-chestwall direction. Thus for scattering correction purposes it is possible to establish a relationship between location breast diameters and local scattering profiles, namely the fitted smooth curves of scatter-to-primary ratios (SPR) and normalized scattered radiations (NSR). In clinical CBCTBI studies, after the local breast diameters are sampled and measured on projection images, the scattering image for every projection image can be generated based on the established relationship, and the projection images can be corrected using either the SPR based method or the NSR based method. Phantom studies and clinical studies showed that both the SPR and NSR methods are able to correct cupping artifacts and reduce reconstruction error. The SPR method does not increase tissue contrast or noise while the NSR method increases both.
引用
收藏
页码:91 / 109
页数:19
相关论文
共 50 条
[21]   Noise suppression in scatter correction for cone-beam CT [J].
Zhu, Lei ;
Wang, Jing ;
Xing, Lei .
MEDICAL PHYSICS, 2009, 36 (03) :741-752
[22]   Simplified method of scatter correction using a beam-stop-array algorithm for cone-beam computed tomography breast imaging [J].
Cai, Weixing ;
Ning, Ruola ;
Conover, David .
OPTICAL ENGINEERING, 2008, 47 (09)
[23]   Algorithm for x-ray beam hardening and scatter correction in low-dose cone-beam CT: Phantom studies [J].
Liu, Wenlei ;
Rong, Junyan ;
Gao, Peng ;
Liao, Qimei ;
Lu, HongBing .
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
[24]   Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies [J].
Dong, Xue ;
Petrongolo, Michael ;
Niu, Tianye ;
Zhu, Lei .
COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2013, 2013
[25]   Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations [J].
Becker, Amy E. ;
Hernandez, Andrew M. ;
Schwoebel, Paul R. ;
Boone, John M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (23)
[26]   Single-scan scatter correction for cone-beam CT using a stationary beam blocker: a preliminary study [J].
Niu, Tianye ;
Zhu, Lei .
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
[27]   Iodine contrast cone beam CT imaging of breast cancer [J].
Partain, Larry ;
Prionas, Stavros ;
Seppi, Edward ;
Virshup, Gary ;
Roos, Gerhard ;
Sutherland, Robert ;
Boone, John .
MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
[28]   Real time flat panel detector-based cone beam volume CT breast imaging: phantom and specimen study [J].
Ning, R ;
Conover, D ;
Yu, Y ;
Schiffhauer, L ;
Lu, XH ;
Cullinan, J ;
Robinson, A .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :181-190
[29]   Evaluation of flat panel detector cone beam CT breast imaging with different sizes of breast phantoms [J].
Ning, R ;
Conover, D ;
Lu, XH ;
Zhang, Y ;
Yu, Y ;
Schiffhauer, L ;
Cullinan, J .
MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 :626-636
[30]   A sinogram based technique for image correction and removal of metal clip artifacts in cone beam breast CT [J].
Wang, T. ;
Shen, Y. ;
Zhong, Y. ;
Lai, C. -J. ;
Wang, J. ;
Shaw, C. C. .
MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033