Minimally invasive methods for the treatment of cancers, such as high-intensity focused ultrasound (HIFU) and high-energy shock waves (SW), have been proposed recently. Their feasibility for treatment of human cancer needs to be confirmed. A simplified model of isolated perfused pig liver that is close to the human liver in vivo has been proposed. The objective was to study the feasibility of deep focused tissue ablation with HIFU and SW in large organs approaching the size of the human liver, The model was demonstrated to be physiologically valid during the first 2 h of anoxic perfusion with a composite saline solution; arterial and portal pressure, enzymes, urea levels and bile secretion remained stable, lit can simulate the major effects of perfusion and physical phenomena that occur in vivo during treatment, Histological analysis revealed no major changes, Previous results obtained in vivo in animal models at a depth of 2-3 cm were successfully reproduced and deeper lesion arrays at 4, 6, 8 and 9 cm from the surface were produced using the same principles. The depth of 9 cm from the liver surface is consistent with an extracorporeal treatment of most of the liver segments in man. Other applications of the model are proposed, particularly for the study of the role of interferences such as ribs and intestinal gas, blood perfusion and respiratory movements. (C) 1998 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1033 / 1043
页数:11
相关论文
共 36 条
  • [1] THE EVALUATION OF THE ISOLATED-PERFUSED LIVER AS A MODEL FOR THE ASSESSMENT OF LIVER PRESERVATION
    BELL, R
    SHIEL, AGR
    DOLAN, P
    MEARS, DC
    WOODMAN, K
    [J]. AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 1993, 63 (01): : 44 - 52
  • [2] EFFECTS OF PHYSICAL PARAMETERS ON HIGH-TEMPERATURE ULTRASOUND HYPERTHERMIA
    BILLARD, BE
    HYNYNEN, K
    ROEMER, RB
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (04) : 409 - 420
  • [3] ELECTRONIC BEAM-STEERING OF SHOCK-WAVES
    CATHIGNOL, D
    BIRER, A
    NACHEF, S
    CHAPELON, JY
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (03) : 365 - 377
  • [4] EFFECT OF BLOOD PERFUSION ON THE ABLATION OF LIVER PARENCHYMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    CHEN, L
    TERHAAR, G
    HILL, CR
    DWORKIN, M
    CARNOCHAN, P
    YOUNG, H
    BENSTED, JPM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (11) : 1661 - 1673
  • [5] COLEMAN DJ, 1985, OPHTHALMOLOGY, V92, P339
  • [6] TREATMENT OF THE DUNNING PROSTATE RAT-TUMOR R3327-AT1 WITH PULSED HIGH-ENERGY ULTRASOUND SHOCK-WAVES (PHEUS) - GROWTH DELAY AND HISTOMORPHOLOGIC CHANGES
    DEBUS, J
    PESCHKE, P
    HAHN, EW
    LORENZ, WJ
    LORENZ, A
    IFFLAENDER, H
    ZABEL, HJ
    VANKAICK, G
    PFEILER, M
    [J]. JOURNAL OF UROLOGY, 1991, 146 (04) : 1143 - 1146
  • [7] THE EFFECTS OF TISSUE HETEROGENEITIES AND LARGE BLOOD-VESSELS ON THE THERMAL EXPOSURE INDUCED BY SHORT HIGH-POWER ULTRASOUND PULSES
    DORR, LN
    HYNYNEN, K
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (01) : 45 - 59
  • [8] Reduction of the grating lobes of annular arrays used in focused ultrasound surgery
    Dupenloup, F
    Chapelon, JY
    Cathignol, DJ
    Sapozhnikov, OA
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) : 991 - 998
  • [9] ESTIMATION OF THE TEMPERATURE INCREASE IN THE FOCUS OF A LITHOTRIPTER FOR THE CASE OF HIGH-RATE ADMINISTRATION
    FILIPCZYNSKI, L
    PIECHOCKI, M
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (02) : 149 - 156
  • [10] FOSTER RS, 1993, EUR UROL, V23, P29