A high-temperature rf SQUID system for magnetocardiography

被引:18
作者
Zeng, XH [1 ]
Soltner, H
Selbig, D
Bode, M
Bick, M
Ruders, F
Schubert, J
Zander, W
Banzet, M
Zhang, Y
Bousack, H
Braginski, AI
机构
[1] Forschungszentrum Julich, Inst Schicht & Ionentech, D-52425 Julich, Germany
[2] Forschungszentrum Julich, Zentral Labor Elekt, D-52425 Julich, Germany
关键词
magnetic field strength; superconducting quantum interference device; biomagnetism;
D O I
10.1088/0957-0233/9/9/033
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A first-order axial electronic gradiometer having a baseline of 10 cm was constructed by assembling two YBa2Cu3O7-delta rf SQUID magnetometers with coplanar tank resonators, each having a white magnetic field resolution of about 20 fT HZ(-1/2) at 77 K. The gradiometer's near-field resolution was about 35 fT HZ(-1/2), including the Dewar flask's noise. A peak-to-peak noise level of 3 pT was obtained in the bandwidth 0.016-250 Hz. Magnetocardiographic (MCG) measurements were performed using this bandwidth. Measurements on human subjects have been conducted in a magnetically shielded room of moderate shielding factor. Using the signal either of the lower magnetometer or of the gradiometer, high-quality heart signal traces could be collected, which were suitable for diagnostic use. A team of physicians, assisted by two of the authors, used the equipment over 10 months to perform MCG measurements in a medical study of about 80 clinical patients with cardiac arrhythmia problems and healthy persons. The system's performance was stable over that whole period.
引用
收藏
页码:1600 / 1608
页数:9
相关论文
共 24 条
[1]   Theory of RF SQUIDs operating in the presence of large thermal fluctuations [J].
Chesca, B .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1998, 110 (5-6) :963-1001
[2]   LOW-NOISE YBA2CU3O7-X-SRTIO3-YBA2CU3O7-X MULTILAYERS FOR IMPROVED SUPERCONDUCTING MAGNETOMETERS (VOL 66, PG 373, 1995) [J].
DANTSKER, E ;
LUDWIG, F ;
KLEINER, R ;
CLARKE, J ;
TEEPE, M ;
LEE, LP ;
ALFORD, NM ;
BUTTON, T .
APPLIED PHYSICS LETTERS, 1995, 67 (05) :725-726
[3]   The development of a high-T-c magnetocardiography system for unshielded environment [J].
David, B ;
Dossel, O ;
Doormann, V ;
Eckart, R ;
Hoppe, W ;
Kruger, J ;
Laudan, H ;
Rabe, G .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :3267-3270
[4]  
DAVIDSON CR, 1996, UNPUB P BIOMAG 96 SA
[5]   Integrated YBa2Cu3O7-x magnetometer for biomagnetic measurements [J].
Drung, D ;
Ludwig, F ;
Muller, W ;
Steinhoff, U ;
Trahms, L ;
Koch, H ;
Shen, YQ ;
Jensen, MB ;
Vase, P ;
Holst, T ;
Freltoft, T ;
Curio, G .
APPLIED PHYSICS LETTERS, 1996, 68 (10) :1421-1423
[6]   INTEGRATED DC SQUID MAGNETOMETER WITH HIGH DV/DB [J].
DRUNG, D ;
CANTOR, R ;
PETERS, M ;
RYHANEN, T ;
KOCH, H .
IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) :3001-3004
[7]   Low noise YBa2Cu3O7-x SQUID magnetometers operated with additional positive feedback [J].
Drung, D ;
Dantsker, E ;
Ludwig, F ;
Koch, H ;
Kleiner, R ;
Clarke, J ;
Krey, S ;
Reimer, D ;
David, B ;
Doessel, O .
APPLIED PHYSICS LETTERS, 1996, 68 (13) :1856-1858
[8]  
ERNE SN, 1981, BIOMAGNETISM, P79
[9]   Estimation of inductance for high-T-c dc-SQUID's in coplanar waveguide [J].
Fuke, H ;
Saitoh, K ;
Utagawa, T ;
Enomoto, Y .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS & EXPRESS LETTERS, 1996, 35 (12A) :L1582-L1584
[10]   A multi-channel high-T-c SQUID system and its application [J].
Itozaki, H ;
Tanaka, S ;
Toyoda, H ;
Hirano, T ;
Haruta, Y ;
Nomura, M ;
Saijou, T ;
Kado, H .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1996, 9 (4A) :A38-A41