Calibration of CCD sensors mounted on geodetic measuring systems

被引:4
作者
Volgyesi, Lajos [1 ]
Toth, Gyula [1 ]
机构
[1] Budapest Univ Technol & Econ, Dept Geodesy & Surveying, Budapest, Hungary
关键词
CCD sensor; Total station; Collimator; Calibration Matrix; Temperature dependence; Astrogeodetic measurements; QDaedalus;
D O I
10.1080/00396265.2019.1703506
中图分类号
TU [建筑科学];
学科分类号
0813 ;
摘要
If the eyepiece of a total station is replaced with a CCD sensor, the most important task is to calibrate the instrument. Calibration establishes a connection between the readings on the horizontal and vertical circles of the total station and the readings in the coordinate system of the CCD sensor. Our calibration method uses a collimator in a few meters distance, which serves as a target at infinity during the calibration. We tested the calibration using a QDaedalus astrogeodetic measuring system. The optimal number of calibration measurements, the optimal raster size, and the temperature dependence of the measurements were investigated.