Radiation dose reference card for interventional radiology procedures: Experience in a tertiary referral centre

被引:4
作者
Varghese, Anna [1 ]
Keshava, Shyamkumar N. [1 ]
Moses, Vinu [1 ]
Koshy, George [1 ]
Mammen, Suraj [1 ]
Ahmed, Munawwar [1 ]
Livingstone, Roshan S. [1 ]
机构
[1] Christian Med Coll & Hosp, Dept Radiol, Vellore, Tamil Nadu, India
关键词
Flat detector; interventional radiology; radiation dose reference card; FLAT-PANEL DETECTOR; INFORMED-CONSENT; EXPOSURE; RISK; CATHETERIZATION; REDUCTION; ORGAN; TIME;
D O I
10.4103/ijri.IJRI_35_19
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Fluoroscopy-guided interventions can potentially increase radiation risk to patients, if awareness on angiographic imaging technique and radiation dose is neglected. Aim: To develop patient radiation dose reference card from standardized imaging techniques for various radiology interventions performed using flat detector based angiography system. Materials and Methods: Real-time monitoring of angiographic exposure parameters and radiation dose were performed for 16 types of radiological interventions. Effective dose (ED) was estimated from dose area product (DAP) using PCXMC Monte Carlo simulation software. Radiation risk levels were estimated based on Biological Effects of Ionising radiation (BEIR) report VII predictive models for an Asian population. Results: Pulse rates of 7.5 pps and 0.6 mm Copper filtration during fluoroscopy and 4 frames per second (fps) and 0.1-0.3 mm Cu filtration during image acquisitions were found to reduce radiation dose. Owing to increased number of image acquisitions, DAP was highest during diagnostic spinal angiography 186.7 Gycm(2) (44.0-377.5). This resulted in highest ED of 59.4 mSv with moderate risk levels (1 in 1000 to 1 in 500). Most of the radiological interventions had low radiation risk levels (1 in 10,000 to 1 in 1000). Conclusion: The patient radiation dose reference card is valuable to the medical community and can aid in patient counselling on radiation induced risk from radiological interventions.
引用
收藏
页码:247 / 252
页数:6
相关论文
共 33 条
  • [1] [Anonymous], 2006, HLTH RISKS EXP LOW L
  • [2] [Anonymous], 1997, 126 NCRP
  • [3] Risk of cancer associated with cardiac catheterization procedures during childhood: a cohort study in France
    Baysson, Helene
    Rehel, Jean Luc
    Boudjemline, Younes
    Petit, Jerome
    Girodon, Brigitte
    Aubert, Bernard
    Laurier, Dominique
    Bonnet, Damien
    Bernier, Marie-Odile
    [J]. BMC PUBLIC HEALTH, 2013, 13
  • [4] Effective dose: a flawed concept that could and should be replaced
    Brenner, D. J.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (967) : 521 - 523
  • [5] Effective dose: A useful concept in diagnostic radiology
    Brenner, David
    Huda, Walter
    [J]. RADIATION PROTECTION DOSIMETRY, 2008, 128 (04) : 503 - 508
  • [6] Effective-dose estimation in interventional radiological procedures
    Falco M.D.
    Masala S.
    Stefanini M.
    Bagalà P.
    Morosetti D.
    Calabria E.
    Tonnetti A.
    Verona-Rinati G.
    [J]. Radiological Physics and Technology, 2018, 11 (2) : 149 - 155
  • [7] Effect of Modern Dose-Reduction Technology on the Exposure of Interventional Cardiologists to Radiation in the Catheterization Laboratory
    Faroux, Laurent
    Blanpain, Thierry
    Nazeyrollas, Pierre
    Tassan-Mangina, Sophie
    Heroguelle, Virginie
    Tourneux, Christophe
    Baudin, Florian
    Metz, Damien
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2018, 11 (02) : 222 - 223
  • [8] Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures
    Fetterly, Kenneth A.
    Lennon, Ryan J.
    Bell, Malcolm R.
    Holmes, David R., Jr.
    Rihal, Charanjit S.
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (03) : 336 - 343
  • [9] Frush DP, 2010, EVIDENCE-BASED IMAGING IN PEDIATRICS: OPTIMIZING IMAGING IN PEDIATRIC PATIENT CARE, P25, DOI 10.1007/978-1-4419-0922-0_3
  • [10] Patient radiation doses and reference levels in pediatric interventional radiology
    Geryes, Bouchra Habib
    Bak, Adeline
    Lachaux, Julie
    Ozanne, Augustin
    Boddaert, Nathalie
    Brunelle, Francis
    Naggara, Olivier
    Saliou, Guillaume