Quality control of viewing boxes at the University Hospital of the Universidade Federal do Rio de Janeiro

被引:0
作者
Azevedo, ACP
Gonçalves, OD
Magalhaes, LAG
Silva, MO
Oliveira, SR
机构
[1] Univ Fed Rio de Janeiro, Dept Radiol, Fac Med, Hosp Clementino Fraga Filho, BR-21941590 Rio De Janeiro, Brazil
[2] Univ Fed Rio de Janeiro, Inst Fis, BR-21941590 Rio De Janeiro, Brazil
[3] Univ Fed Rio de Janeiro, Dept Radiol, Fac Med, BR-21941590 Rio De Janeiro, Brazil
关键词
viewing box; diagnostic radiology; quality control; luminance levels;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
At the University Hospital of the Universidade Federal do Rio de Janeiro, a survey was performed to analyze the working conditions of viewing boxes. The survey consisted in checking the illuminance of the rooms, the luminance levels of the viewing boxes and their uniformity. Results show that the luminance levels are all below the recommended levels, that the uniformity is generally inadequate, and that the room illuminance is adequate in only 19% of the rooms when compared to recommendations contained in international quality control protocols. Cleaning and minor repairs showed that an improvement of around 50% in luminance levels can be achieved at extremely low cost.
引用
收藏
页码:89 / 92
页数:4
相关论文
共 10 条
[1]  
Azevedo A. C. P., 1999, RADIOL BRAS, V32, P309
[2]  
BAULM A, 1982, WHO TRAIN WORKSH QUA
[3]  
*BRIT I RAD, 1988, ASS QUAL DIAGN
[4]  
*CEC, 1989, XII17390 CEC
[5]   QUALITY ASSURANCE OF VIEWING BOXES - PROPOSAL FOR ESTABLISHING MINIMUM REQUIREMENTS AND RESULTS FROM A SPANISH QUALITY-CONTROL PROGRAM [J].
GUIBELALDE, E ;
VANO, E ;
LLORCA, AL .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (751) :564-567
[6]  
HARTMANN E, 1989, BIR REPORT, V18, P135
[7]   Film viewing conditions in mammography [J].
Hill, SJ ;
Faulkner, K ;
Law, J ;
Starritt, HC .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832) :409-411
[8]  
MOORES BM, 1987, PRACTICAL GUIDE QUAL
[9]  
*NEMA, 1980, NEMA PUBL
[10]  
WHO, 1982, QUAL ASS DIAGN RAD