A Systematic Review on Complications of Tissue Preservation Surgical Techniques in Percutaneous Bone Conduction Hearing Devices

被引:50
|
作者
Verheij, Emmy [1 ,2 ]
Bezdjian, Aren [1 ,2 ]
Grolman, Wilko [1 ,2 ]
Thomeer, Henricus G. X. M. [1 ,2 ]
机构
[1] Univ Med Ctr Utrecht, Dept Otorhinolaryngol Head & Neck Surg, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Brain Ctr Rudolf Magnus, Utrecht, Netherlands
关键词
Bone conduction; Complications; Holgers classification; Operative time; Percutaneous; SURGERY RANDOMIZED-TRIAL; IMPLANT-SURGERY; LINEAR INCISION; SKIN REACTIONS; AIDS; REDUCTION; DERMATOME; CHILDREN;
D O I
10.1097/MAO.0000000000001091
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To investigate skin-related postoperative complications from tissue preservation approaches in percutaneous bone conduction device (BCD) implantations. Data Sources: PubMed, Embase, and Cochrane Library. Study Selection: We identified studies on BCDs including the opted surgical technique and derived complications. Retrieved articles were screened using predefined inclusion criteria. Critical appraisal included directness of evidence and risk of bias. Studies that successfully passed critical appraisal were included. Data Extraction: Outcome measures included patient demographics, surgery time, follow-up time, and complications reported by Holgers classification. Data Synthesis: We selected 18 articles for data extraction; encompassing 381 BCDs implanted using non-skin thinning approaches. Four studies reported an implantation technique using the punch method (81 implants), 13 studies applied the linear incision technique without soft tissue reduction (288 implants) and one study used the Weber technique (12 implants). Holgers 3 was described in 2.5% following the punch technique, in 5.9% following the linear incision technique, and in no implants following the Weber technique. One patient was mentioned having Holgers 4, skin overgrowth was reported in 6 patients, and 10 studies compared their non-skin thinning technique with a skin-thinning technique. Overall, the soft tissue preservation technique had a similar or superior complication rate, shorter surgical time, and better and faster healing, compared with the soft tissue reduction technique. Conclusion: Tissue preservation surgical techniques for percutaneous BCDs have limited postoperative skin complication rates. Moreover, these techniques are suggested to have at least similar complications rates compared with skin-thinning techniques.
引用
收藏
页码:829 / 837
页数:9
相关论文
共 50 条
  • [21] The role of bone conduction hearing aids in congenital unilateral hearing loss: A systematic review
    Liu, C. Carrie
    Livingstone, Devon
    Yunker, Warren K.
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2017, 94 : 45 - 51
  • [22] Modern Advances in Bone Conduction–Hearing Devices
    Geoffrey C. Casazza
    Bradley W. Kesser
    Current Otorhinolaryngology Reports, 2022, 10 : 370 - 376
  • [23] Bone Conduction Benefits and Limitations of Surgical and Nonsurgical Devices
    Snapp, Hillary
    OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2021, 54 (06) : 1205 - 1217
  • [24] Ex vivo Evaluation of a New Drill System for Placement of Percutaneous Bone Conduction Devices
    Strijbos, Ruben M.
    Straatman, Louise V.
    Stokroos, Robert J.
    Johansson, Martin L.
    FRONTIERS IN SURGERY, 2022, 9
  • [25] Systematic Review on Surgical Outcomes and Hearing Preservation for Cochlear Implantation in Children and Adults
    Bruijnzeel, Hanneke
    Draaisma, Kaspar
    van Grootel, Roderick
    Stegeman, Inge
    Topsakal, Vedat
    Grolman, Wilko
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2016, 154 (04) : 586 - 596
  • [26] Fitting bone conduction hearing devices to children: audiological practices and challenges
    Gordey, Dave
    Bagatto, Marlene
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2021, 60 (05) : 385 - 392
  • [27] Predisposing factors for adverse skin reactions with percutaneous bone anchored hearing devices implanted with skin reduction techniques
    Candreia, Claudia
    Birrer, Ruth
    Fistarol, Susanna
    Kompis, Martin
    Caversaccio, Marco
    Arnold, Andreas
    Stieger, Christof
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2016, 273 (12) : 4185 - 4192
  • [28] Clinical Outcomes of Soft Tissue Preservation Surgery With Hydroxyapatite-Coated Abutments Compared to Traditional Percutaneous Bone Conduction Hearing Implant Surgery-A Pragmatic Multi-Center Randomized Controlled Trial
    van Hoof, M.
    Wigren, S.
    Blechert, J. Ivarsson
    Joore, M. A.
    Mateijsen, D. J. M.
    Bom, S. J. H.
    Stalfors, J.
    Eeg-Olofsson, Mans
    Deguine, O.
    van der Rijt, A. J. M.
    Flynn, M. C.
    Marco Algarra, J.
    Stokroos, R. J.
    FRONTIERS IN SURGERY, 2020, 7
  • [29] Air Conduction, Bone Conduction, and Soft Tissue Conduction Audiograms in Normal Hearing and Simulated Hearing Losses
    Adelman, Cahtia
    Cohen, Adi
    Regev-Cohen, Adi
    Chordekar, Shai
    Fraenkel, Rachel
    Sohmer, Haim
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2015, 26 (01) : 101 - 108
  • [30] Long-Term Outcomes of the Minimally Invasive Ponto Surgery vs. Linear Incision Technique With Soft Tissue Preservation for Installation of Percutaneous Bone Conduction Devices
    Strijbos, Ruben M.
    Straatman, Louise, V
    Calon, Tim G. A.
    Johansson, Martin L.
    de Bruijn, Arthur J. G.
    van den Berge, Herbert
    Wagenaar, Mariette
    Eichhorn, Edwin
    Janssen, Miranda
    Jonhede, Sofia
    van Tongeren, Joost
    Holmberg, Marcus
    Stokroos, Robert
    FRONTIERS IN NEUROLOGY, 2021, 12