Design and application of an assessment protocol for electromagnetic tracking systems

被引:106
作者
Hummel, JB [1 ]
Bax, MR
Figl, ML
Kang, Y
Maurer, C
Birkfellner, WW
Bergmann, H
Shahidi, R
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, Vienna, Austria
[2] Stanford Univ, Sch Med, Image Guidance Labs, Stanford, CA 94305 USA
[3] Ludwig Boltzmann Inst Nucl Med & Endocrinol, Vienna, Austria
基金
奥地利科学基金会;
关键词
D O I
10.1118/1.1944327
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper defines a simple protocol for competitive and quantified evaluation of electromagnetic tracking systems such as the NDI Aurora (A) and Ascension microBIRD with dipole transmitter (B). It establishes new methods and a new phantom design which assesses the reproducibility and allows comparability with different tracking systems in a consistent environment. A machined base plate was designed and manufactured in which a 50 mm grid of holes was precisely drilled for position measurements. In the center a circle of 32 equispaced holes enables the accurate measurement of rotation. The sensors can be clamped in a small mount which fits into pairs of grid holes on the base plate. Relative positional/orientational errors are found by subtracting the known distances/ rotations between the machined locations from the differences of the mean observed positions/ rotation. To measure the influence of metallic objects we inserted rods made of steel (SST 303, SST 416), aluminum, and bronze into the sensitive volume between sensor and emitter. We calculated the fiducial registration error and fiducial location error with a standard stylus calibration for both tracking systems and assessed two different methods of stylus calibration. The positional jitter amounted to 0.14 mm(A) and 0.08 mm(B). A relative positional error of 0.96 mm +/- 0.68 mm, range -0.06 mm; 2.23 mm(A) and 1.14 mm 0.78 mm, range -3.72 min; 1.57 mm(B) for a given distance of 50 mm was found. The relative rotation error was found to be 0.51 degrees (A)/0.04 degrees (B). The most relevant distortion caused by metallic objects results from SST 416. The maximum error 4.2 mm(A) / >= 100 min(B) occurs when the rod is close to the sensor(20 mm). While (B) is more sensitive with respect to metallic objects, (A) is less accurate concerning orientation measurements. (B) showed a systematic error when distances are calculated. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2371 / 2379
页数:9
相关论文
共 24 条
  • [1] Adelstein B. D., 1992, UIST. Fifth Annual Symposium on User Interface Software and Technology. Proceedings, P15, DOI 10.1145/142621.142624
  • [2] Systematic distortions in magnetic position digitizers
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Enislidis, G
    Kollmann, C
    Rafolt, D
    Nowotny, R
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2242 - 2248
  • [3] Calibration of tracking systems in a surgical environment
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Ewers, R
    Bergmann, H
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 737 - 742
  • [4] De Groot NM, 2000, J CARDIOVASC ELECTR, V11, P1183
  • [5] 3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION
    DETMER, PR
    BASHEIN, G
    HODGES, T
    BEACH, KW
    FILER, EP
    BURNS, DH
    STRANDNESS, DE
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) : 923 - 936
  • [6] Predicting error in rigid-body point-based registration
    Fitzpatrick, JM
    West, JB
    Maurer, CR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 694 - 702
  • [7] Accuracy assessment protocols for electromagnetic tracking systems
    Frantz, DD
    Wiles, AD
    Leis, SE
    Kirsch, SR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14) : 2241 - 2251
  • [8] In vitro evaluation of three-dimensional ultrasonography based on magnetic scanhead tracking
    Gilja, OH
    Hausken, T
    Olafsson, S
    Matre, K
    Odegaard, S
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (08) : 1161 - 1167
  • [9] A technique for validating image-guided gene therapy
    Harris, S
    Hallahan, D
    Galloway, R
    [J]. MEDICAL IMAGING 2003: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2003, 5029 : 330 - 337
  • [10] HARTOV A, 1999, NEUROSURG FOCUS, V6