Complete calibration of a stereo photogrammetric system through control points of unknown coordinates

被引:32
作者
Cerveri, P
Borghese, NA
Pedotti, A
机构
[1] LITA, CNR, Lab Human Mot Study & Virtual Real, Ist Neurosci & Bioimmagini, I-20090 Milan, Italy
[2] Politecn Milan, Dipartimento Bioingn, Ctr Bioingn, Fdn ProJuventute, I-20148 Milan, Italy
关键词
calibration; epipolar geometry; evolution strategies; motion analysis; 3D reconstruction;
D O I
10.1016/S0021-9290(98)00104-3
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper presents a new method for calibrating a video 3D stereo-photogrammetric system. The external parameters and the focal lengths of the cameras are determined from the epipolar constraint and the principal points are computed through the minimisation of a cost function carried out through an evolutionary optimisation. The method has been made more robust with a deterministic annealing procedure of the search region amplitude. Calibration is carried out by moving a rigid bar, carrying two markers on its extremities, inside the working volume. The distance between the two markers is the only measure required. Tests on real data are reported which show that the obtained accuracy is comparable to the one achieved calibrating with control points of known 3D coordinates. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:935 / 940
页数:6
相关论文
共 23 条