Measurement of Gd-DTPA dialysis clearance rates by using a look-locker imaging technique

被引:18
作者
Niendorf, ER
Santyr, GE
Brazy, PC
Grist, TM
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT RADIOL,MADISON,WI 53706
[3] UNIV WISCONSIN,DEPT NEPHROL,MADISON,WI 53706
[4] UNIV WISCONSIN,MD PHD PROGRAM,MADISON,WI 53706
关键词
MRI; Look-Locker; clearance rate; filtration fraction;
D O I
10.1002/mrm.1910360412
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The exponential clearance rate constant, (kappa), and filtration fraction (FF) have been measured for dialysis of Godolinium-diethylenetriaminepentaacetic acid (Gd-DTPA) (Magnevist, Berlex, Wayne, NJ) solutions by using a Look-Locker imaging technique under conditions of flow. The measured values of kappa for the Baxter CA-50, CA-110, and CA-210 filters were 0.0037 +/- 0.0003, 0.0057 +/- 0.0017, and 0.0092 +/- 0.0018 min(-1), respectively, for dialysis of 4.0 liters of aqueous Gd-DTPA solutions, The measured values of FF for the Baxter CA-110 and CA-210 filters were 0.060 +/- 0.013 and 0.089 +/- 0.015, respectively, for dialysis of aqueous Gd-DTPA at 350 ml/min, The kappa and FF measurements agree with values that use inversion recovery (IR) on static samples obtained by drawing aliquots of solution during the course of dialysis, This in vitro experiment suggests that accurate in vivo measurements of filtration fraction and glomerular filtration rate (GFR) may be possible.