Use of the Fully Spectroscopic Pixelated Cadmium Telluride Detector for Benchtop X-Ray Fluorescence Computed Tomography

被引:0
|
作者
Jayarathna, Sandun [1 ]
Moktan, Hem [1 ]
Cho, Sang Hyun [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
来源
IEEE ACCESS | 2022年 / 10卷
关键词
Detectors; Collimators; Imaging; X-ray imaging; Image quality; Energy resolution; Economic indicators; Fluorescence; Nanoparticles; X-ray fluorescence; X-ray fluorescence computed tomography; CdTe detector; HEXITEC; gold nanoparticle; quantitative imaging; XFCT;
D O I
10.1109/ACCESS.2022.3210590
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this work, we integrated a commercially-available fully-spectroscopic pixelated cadmium telluride (CdTe) detector system as a two-dimensional (2D) array detector into our existing benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system. After integrating this detector, known as High-Energy X-ray Imaging Technology (HEXITEC), we performed quantitative imaging of gold nanoparticle (GNP) distribution in a small animal-sized phantom using our benchtop XFCT system. Owing to the upgraded detector component within our benchtop XFCT system, we were able to conduct this phantom imaging in an unprecedented manner by volumetric XFCT scans followed by XFCT image reconstruction in 3D. The current results showed that adoption of HEXITEC, in conjunction with a custom-made parallel-hole collimator, drastically reduced the XFCT scan time/dose. Compared with the previous work performed with our original benchtop XFCT system adopting a single crystal CdTe detector, the currently observed reduction was up to a factor of 5, while achieving comparable GNP detection limit under similar experimental conditions. Overall, we demonstrated, for the first time to the best our knowledge, the feasibility of benchtop XFCT imaging of small animal-sized objects containing biologically relevant GNP concentrations (on the order of 0.1 mg Au/cm3 or 100 parts-per-million/ppm), with the scan time (on the order of 1 minute)/ x-ray dose (on the order of 10 cGy) that are likely meeting the minimum requirements for routine preclinical imaging applications.
引用
收藏
页码:105074 / 105083
页数:10
相关论文
共 50 条
  • [41] Low-dose low-scattering X-ray computed tomography with high-spatial-energy resolutions using a cooled cadmium telluride detector
    Moriyama, Hodaka
    Watanabe, Manabu
    Kusachi, Shinya
    Oda, Yasuyuki
    Sato, Eiichi
    ULTRAMICROSCOPY, 2019, 199 : 62 - 69
  • [42] Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography
    Manohar, Nivedh
    Reynoso, Francisco J.
    Diagaradjane, Parmeswaran
    Krishnan, Sunil
    Cho, Sang Hyun
    SCIENTIFIC REPORTS, 2016, 6
  • [43] Optimization of a table-top x-ray fluorescence computed tomography (XFCT) system
    Dunning, C. A. S.
    Bazalova-Carter, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23)
  • [44] Sheet beam x-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticles
    Dunning, Chelsea A. S.
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2018, 45 (06) : 2572 - 2582
  • [45] Simultaneous fluorescence and Compton scattering computed tomography based on linear polarization X-ray
    Chi, Zhi-Jun
    Zhang, Hong-Ze
    Lin, Jin
    Zhang, Xuan-Qi
    Ding, Hao
    Tian, Qi-Li
    Zhang, Zhi
    Du, Ying-Chao
    Huang, Wen-Hui
    Tang, Chuan-Xiang
    NUCLEAR SCIENCE AND TECHNIQUES, 2024, 35 (10)
  • [46] Optimized Detector Angle for Improving signal-to-noise ratio of Pinhole X-ray Fluorescence Computed Tomography
    Guo, Jing
    Feng, Peng
    Deng, Luzhen
    Luo, Yan
    He, Peng
    ELEVENTH INTERNATIONAL CONFERENCE ON INFORMATION OPTICS AND PHOTONICS (CIOP 2019), 2019, 11209
  • [47] Reduced-scan schemes for x-ray fluorescence computed tomography
    La Riviere, Patrick J.
    Vargas, Phillip
    Newville, Matt
    Sutton, Stephen R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) : 1535 - 1542
  • [48] Optimization of Detection Angle for Pinhole X-Ray Fluorescence Computed Tomography
    Guo Jing
    Feng Peng
    Deng Luzhen
    Luo Yan
    He Peng
    Wei Biao
    ACTA OPTICA SINICA, 2020, 40 (01)
  • [49] Spectrally resolving and scattering-compensated x-ray luminescence/fluorescence computed tomography
    Cong, Wenxiang
    Shen, Haiou
    Wang, Ge
    JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (06)
  • [50] Experimental simulation of a spectroscopic pixel X-ray detector
    Reinhart, C
    Giersch, J
    Uhlmann, N
    Anton, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 546 (1-2) : 205 - 208