Use of the Fully Spectroscopic Pixelated Cadmium Telluride Detector for Benchtop X-Ray Fluorescence Computed Tomography

被引:0
|
作者
Jayarathna, Sandun [1 ]
Moktan, Hem [1 ]
Cho, Sang Hyun [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
来源
IEEE ACCESS | 2022年 / 10卷
关键词
Detectors; Collimators; Imaging; X-ray imaging; Image quality; Energy resolution; Economic indicators; Fluorescence; Nanoparticles; X-ray fluorescence; X-ray fluorescence computed tomography; CdTe detector; HEXITEC; gold nanoparticle; quantitative imaging; XFCT;
D O I
10.1109/ACCESS.2022.3210590
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this work, we integrated a commercially-available fully-spectroscopic pixelated cadmium telluride (CdTe) detector system as a two-dimensional (2D) array detector into our existing benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system. After integrating this detector, known as High-Energy X-ray Imaging Technology (HEXITEC), we performed quantitative imaging of gold nanoparticle (GNP) distribution in a small animal-sized phantom using our benchtop XFCT system. Owing to the upgraded detector component within our benchtop XFCT system, we were able to conduct this phantom imaging in an unprecedented manner by volumetric XFCT scans followed by XFCT image reconstruction in 3D. The current results showed that adoption of HEXITEC, in conjunction with a custom-made parallel-hole collimator, drastically reduced the XFCT scan time/dose. Compared with the previous work performed with our original benchtop XFCT system adopting a single crystal CdTe detector, the currently observed reduction was up to a factor of 5, while achieving comparable GNP detection limit under similar experimental conditions. Overall, we demonstrated, for the first time to the best our knowledge, the feasibility of benchtop XFCT imaging of small animal-sized objects containing biologically relevant GNP concentrations (on the order of 0.1 mg Au/cm3 or 100 parts-per-million/ppm), with the scan time (on the order of 1 minute)/ x-ray dose (on the order of 10 cGy) that are likely meeting the minimum requirements for routine preclinical imaging applications.
引用
收藏
页码:105074 / 105083
页数:10
相关论文
共 50 条
  • [31] An analytical simulation method for X-ray fluorescence computed tomography
    Gao, Bo
    Laforce, Brecht
    Dhaene, Jelle
    Vincze, Laszlo
    Van Hoorebeke, Luc
    Boone, Matthieu N.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 170 (170)
  • [32] Novel sampling strategies for x-ray fluorescence computed tomography
    La Riviere, Patrick J.
    Vargas, Phillip
    DEVELOPMENTS IN X-RAY TOMOGRAPHY VI, 2008, 7078
  • [33] Computed tomography and X-ray fluorescence CT of biological samples
    Pereira, G. R.
    Anjos, M. J.
    Rocha, H. S.
    Faria, P.
    Perez, C. A.
    Lopes, R. T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (02) : 951 - 954
  • [34] Feasibility study of Compton cameras for x-ray fluorescence computed tomography with humans
    Vernekohl, Don
    Ahmad, Moiz
    Chinn, Garry
    Xing, Lei
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (24) : 8521 - 8540
  • [35] Experimental demonstration of novel imaging geometries for x-ray fluorescence computed tomography
    Fu, Geng
    Meng, Ling-Jian
    Eng, Peter
    Newville, Matt
    Vargas, Phillip
    La Riviere, Patrick
    MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Element Mapping in Organic Samples Utilizing a Benchtop X-Ray Fluorescence Emission Tomography (XFET) System
    Groll, A.
    George, J.
    Vargas, P.
    La Riviere, P. J.
    Meng, L. J.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 2310 - 2317
  • [37] Region of interest reconstruction in x-ray fluorescence computed tomography
    La Riviere, Patrick J.
    Vargas, Phillip
    Xia, Dan
    Pan, Xiaochuan
    DEVELOPMENTS IN X-RAY TOMOGRAPHY VI, 2008, 7078
  • [38] X-Ray Luminescence and X-Ray Fluorescence Computed Tomography: New Molecular Imaging Modalities
    Ahmad, Moiz
    Pratx, Guillem
    Bazalova, Magdalena
    Xing, Lei
    IEEE ACCESS, 2014, 2 : 1051 - 1061
  • [39] Investigation of transmission computed tomography (CT) image quality and x-ray dose achievable from an experimental dual-mode benchtop x-ray fluorescence CT and transmission CT system
    Deng, Luzhen
    Yasar, Selcuk
    Ahmed, Md Foiez
    Jayarathna, Sandun
    Feng, Peng
    Wei, Biao
    Vedantham, Srinivasan
    Karellas, Andrew
    Cho, Sang Hyun
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2019, 27 (03) : 431 - 442
  • [40] High-spatial-resolution nanoparticle X-ray fluorescence tomography
    Larsson, Jakob C.
    Vagberg, William
    Vogt, Carmen
    Lundstrom, Ulf
    Larsson, Daniel H.
    Hertz, Hans M.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783