Use of the Fully Spectroscopic Pixelated Cadmium Telluride Detector for Benchtop X-Ray Fluorescence Computed Tomography

被引:0
|
作者
Jayarathna, Sandun [1 ]
Moktan, Hem [1 ]
Cho, Sang Hyun [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
来源
IEEE ACCESS | 2022年 / 10卷
关键词
Detectors; Collimators; Imaging; X-ray imaging; Image quality; Energy resolution; Economic indicators; Fluorescence; Nanoparticles; X-ray fluorescence; X-ray fluorescence computed tomography; CdTe detector; HEXITEC; gold nanoparticle; quantitative imaging; XFCT;
D O I
10.1109/ACCESS.2022.3210590
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this work, we integrated a commercially-available fully-spectroscopic pixelated cadmium telluride (CdTe) detector system as a two-dimensional (2D) array detector into our existing benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system. After integrating this detector, known as High-Energy X-ray Imaging Technology (HEXITEC), we performed quantitative imaging of gold nanoparticle (GNP) distribution in a small animal-sized phantom using our benchtop XFCT system. Owing to the upgraded detector component within our benchtop XFCT system, we were able to conduct this phantom imaging in an unprecedented manner by volumetric XFCT scans followed by XFCT image reconstruction in 3D. The current results showed that adoption of HEXITEC, in conjunction with a custom-made parallel-hole collimator, drastically reduced the XFCT scan time/dose. Compared with the previous work performed with our original benchtop XFCT system adopting a single crystal CdTe detector, the currently observed reduction was up to a factor of 5, while achieving comparable GNP detection limit under similar experimental conditions. Overall, we demonstrated, for the first time to the best our knowledge, the feasibility of benchtop XFCT imaging of small animal-sized objects containing biologically relevant GNP concentrations (on the order of 0.1 mg Au/cm3 or 100 parts-per-million/ppm), with the scan time (on the order of 1 minute)/ x-ray dose (on the order of 10 cGy) that are likely meeting the minimum requirements for routine preclinical imaging applications.
引用
收藏
页码:105074 / 105083
页数:10
相关论文
共 50 条
  • [21] Energy discriminating x-ray camera utilizing a cadmium telluride detector
    Sato, Eiichi
    Purkhet, Abderyim
    Matsukiyo, Hiroshi
    Osawa, Akihiro
    Enomoto, Toshiyuki
    Wantanabe, Manabu
    Nagao, Jiro
    Nomiya, Seiichiro
    Hitomi, Keitaro
    Tanaka, Etsuro
    Kawai, Toshiaki
    Sato, Shigehiro
    Ogawa, Akira
    Onagawa, Jun
    OPTICAL ENGINEERING, 2009, 48 (07)
  • [22] Monte Carlo study of x-ray detection configurations for benchtop x-ray fluorescence computed tomography of gold nanoparticle-loaded objects
    Moktan, Hem
    Ahmed, Md Foiez
    Jayarathna, Sandun
    Deng, Luzhen
    Cho, Sang Hyun
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17)
  • [23] Quantitative Imaging of Gd Nanoparticles in Mice Using Benchtop Cone-Beam X-ray Fluorescence Computed Tomography System
    Zhang, Siyuan
    Li, Liang
    Chen, Jiayou
    Chen, Zhiqiang
    Zhang, Wenli
    Lu, Hongbing
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2019, 20 (09)
  • [24] Scattering Noise Model Enhanced EM-TV Algorithm for Benchtop X-ray Fluorescence Computed Tomography Image Reconstruction
    Zhang, Siyuan
    Li, Liang
    Chen, Zhiqiang
    IEEE ACCESS, 2019, 7 : 113589 - 113595
  • [25] The progress of X-ray fluorescence computed tomography at SSRF
    Deng, Biao
    Yang, Qun
    Du, Guohao
    Tong, Yajun
    Xie, Honglan
    Xiao, Tiqiao
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 305 : 5 - 8
  • [26] X-ray Fluorescence Computed Tomography (XFCT) Imaging with a Superfine Pencil Beam X-ray Source
    Romero, Ignacio O.
    Fang, Yile
    Lun, Michael
    Li, Changqing
    PHOTONICS, 2021, 8 (07)
  • [27] Energy-discriminating x-ray camera utilizing a cadmium telluride detector
    Sato, Eiichi
    Abderyim, Purkhet
    Matsukiyo, Hiroshi
    Osawa, Akihiro
    Enomoto, Toshiyuki
    Watanabe, Manabu
    Nagao, Jiro
    Nomiya, Seichiro
    Hitomi, Keitaro
    Ogawa, Akira
    Sato, Shigehiro
    Ichimaru, Toshio
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS X, 2008, 7079
  • [28] Feasibility Simulation of 3D Benchtop Multi-Pinhole X-ray Fluorescence Computed Tomography with Two Novel Geometries
    Ye, Binqiang
    Deng, Luzhen
    Jiang, Shanghai
    Cao, Sijun
    Zhao, Ruge
    Feng, Peng
    PHOTONICS, 2023, 10 (04)
  • [29] Enabling Low-Dose In Vivo Benchtop X-ray Fluorescence Computed Tomography through Deep-Learning-Based Denoising
    Mahmoodian, Naghmeh
    Rezapourian, Mohammad
    Inamdar, Asim Abdulsamad
    Kumar, Kunal
    Fachet, Melanie
    Hoeschen, Christoph
    JOURNAL OF IMAGING, 2024, 10 (06)
  • [30] Polarized x-ray excitation for scatter reduction in x-ray fluorescence computed tomography
    Vernekohl, Don
    Tzoumas, Stratis
    Zhao, Wei
    Xing, Lei
    MEDICAL PHYSICS, 2018, 45 (08) : 3741 - 3748