Direct cadmium determination in sediment samples by flame atomic absorption spectrometry using multivariate calibration procedures

被引:16
作者
Flores, EMD [1 ]
Paniz, JNG
Saidelles, APF
Müller, EI
da Costa, AB
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
[2] Univ Santa Cruz, Dept Quim & Fis, BR-96815900 Santa Cruz Do Sul, RS, Brazil
[3] Univ Santa Cruz, Dept Biol, BR-96815900 Santa Cruz Do Sul, RS, Brazil
关键词
D O I
10.1039/b212455f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A direct solid sampling procedure by flame atomic absorption spectrometry (DSA-FAAS) has been developed for the determination of cadmium in marine and river sediment samples. Multivariate calibration using the partial least squares (PLS) regression procedure was applied to calibration using dry sediment reference samples. Test-samples were introduced into the flame as a dry aerosol using a T-quartz cell set between the burner and the optical path. Background signals were always low and a characteristic mass of 0.21 ng Cd was found and an absolute limit of detection of 0.25 ng (3s), or 0.25 mug g(-1) if a sample mass of 1 mg was used. Using the proposed multivariate calibration procedure the results were considered satisfactory regarding both agreement with the results from reference samples (from 96.8 to 101.7%) and precision ( relative standard deviation from 5.5 to 7.0%). With this procedure errors occurring when conventional univariate calibration was used (up to 64%) were reduced to less than 4%.