Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: Epicardial bipolar and endocardial bipolar sensing circuits are compared

被引:20
作者
Glotzer, TV [1 ]
Gordon, M [1 ]
Sparta, M [1 ]
Radoslovich, G [1 ]
Zimmerman, J [1 ]
机构
[1] Hackensack Univ, Med Ctr, Dept Med, Div Cardiol,Sect Electrophysiol, Hackensack, NJ 07601 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 10期
关键词
electrical muscle stimulation; implantable cardioverter defibrillator; electromagnetic interference;
D O I
10.1111/j.1540-8159.1998.tb00021.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This case report is about two patients with live different types of ICDs who underwent electrical muscle stimulation (EMS) therapy. In one patient with an ICD that has epicardial screw-in bipolar sensing leads, electromagnetic interference (EMI) from the EMS device caused the delivery of an inappropriate ICD discharge. In a second patient with an ICD with endocardial true bipolar sensing, there was no evidence of EMI during the EMS therapy despite all of our attempts to reproduce if. The sensing circuits in the two different ICDs are compared.
引用
收藏
页码:1996 / 1998
页数:3
相关论文
共 12 条
  • [1] INADVERTENT AICD INACTIVATION WHILE PLAYING BINGO
    FERRICK, KJ
    JOHNSTON, D
    KIM, SG
    ROTH, J
    BRODMAN, R
    ZIMMERMAN, J
    FISHER, JD
    [J]. AMERICAN HEART JOURNAL, 1991, 121 (01) : 206 - 207
  • [2] KARSON TH, 1989, NEW ENGL J MED, V320, P1628
  • [3] INTERFERENCE FROM A HAND-HELD RADIOFREQUENCY REMOTE-CONTROL CAUSING DISCHARGE OF AN IMPLANTABLE DEFIBRILLATOR
    MAN, KC
    DAVIDSON, T
    LANGBERG, JJ
    MORADY, F
    KALBFLEISCH, SJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (08): : 1756 - 1758
  • [4] Environmental electromagnetic interference from electronic article surveillance devices: Interactions with an ICD
    McIvor, ME
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (12): : 2229 - 2230
  • [5] CAN TRANS-CUTANEOUS ELECTRICAL NERVE-STIMULATION BE SAFELY USED IN PATIENTS WITH PERMANENT CARDIAC-PACEMAKERS
    RASMUSSEN, MJ
    HAYES, DL
    VLIETSTRA, RE
    THORSTEINSSON, G
    [J]. MAYO CLINIC PROCEEDINGS, 1988, 63 (05) : 443 - 445
  • [6] SCHMITT C, 1991, PACE, V14, P982
  • [7] ERRONEOUS DISCHARGE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR CAUSED BY AN ELECTRIC RAZOR
    SEIFERT, T
    BLOCK, M
    BORGGREFE, M
    BREITHARDT, G
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08): : 1592 - 1594
  • [8] NORMALLY FUNCTIONING INHIBITED PACEMAKER CONCOMITANT WITH VERTEBRAL NERVE STIMULATOR
    SYLVEN, C
    LEVANDERLINDGREN, M
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (06): : 645 - 647
  • [9] TECHNICAL MANUAL TEL
  • [10] INSTRUCTION MANUAL E