Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms: Strategies in stent deployment and midterm follow-up

被引:309
作者
Biondi, Alessandra
Janardhan, Vallabh
Katz, Jeffrey M.
Salvaggio, Kimberly
Riina, Howard A.
Gobin, Y. Pierre
机构
[1] Cornell Univ, New York Presbyterian Hosp, Weill Med Coll, Div Intervent Neuroradiol, Ithaca, NY 10021 USA
[2] Cornell Univ, New York Presbyterian Hosp, Weill Med Coll, Dept Neurosurg,Div Intervent Neuroradiol, Ithaca, NY 14853 USA
关键词
aneurysm; coils; endovascular; intracranial; neuroform; stent;
D O I
10.1227/01.NEU.0000290890.62201.A9
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE:To evaluate the midterm results of intracranial stent-assisted coil embolization in the treatment of wide-necked cerebral aneurysms and to assess the efficacy of various strategies used in stent deployment. METHODS: A retrospective study of 42 patients with 46 wide-necked cerebral aneurysms enrolled in a prospective single-center registry of patients treated with a Neuroform stent (Boston Scientific/Target, Fremont, CA), a flexible self-expanding nitinol stent, was performed. Twenty-seven of 46 aneurysms were unruptured aneurysms, 14 were recanalized aneurysms, and five were acutely ruptured. Thirty-nine aneurysms were located in the anterior and seven in the posterior circulation. Mean aneurysm size was 9.8 mm. Stenting before coiling was performed in 13 of 45 aneurysms (29%), coiling before stenting in 27 of 45 aneurysms (60%), and stenting alone in five of 45 aneurysms (11%). The balloon remodeling technique for coiling before stenting was performed in 77% of patients. Angiographic and clinical follow up was available. in 31 patients with 33 aneurysms and ranged from 3 to 24 months. RESULTS: Neuroform stenting was attempted in 46 wide-necked aneurysms (42 patients). Forty-nine stent sessions were performed, including three poststent retreatments. In 46 of 49 sessions (94%), successful deployment of 47 stents for 45 aneurysms was obtained. In 40 aneurysms treated with stent-assisted coiling, angiographic results showed 14 (35%) aneurysm occlusions, 18 (45%) neck remnants, and eight (20%) residual aneurysms. In five recanalized aneurysms treated with stenting alone, no changes were observed in four (80%) aneurysms and one (20%) neck remnant reduced in size. At angiographic follow-up in 30 aneurysms treated with stent-assisted coiling, there were 17 (57%) aneurysm occlusions, seven (23%) neck remnants, and six (20%) residual aneurysms. In three recanalized aneurysms treated with stent alone, two (67%) neck remnants remained unchanged and one (33%) neck remnant decreased in size. Procedural morbidity was observed in two of 42 patients (4.8%) and one patient died. On clinical follow-up, the modified Rankin Scale score was 0 in 27 patients (87%), 1 in three patients (10%), and 2 (3%) in one patient. No aneurysm bled during the follow-up period. CONCLUSION: These results indicate that Neuroform stent-assisted coil embolization is a safe and effective technique in the treatment of wide-necked cerebral aneurysms. Further studies are needed to evaluate the long-term durability of stent-assisted aneurysm occlusion and tolerance to the stent.
引用
收藏
页码:460 / 468
页数:9
相关论文
共 29 条
  • [1] Akpek S, 2005, AM J NEURORADIOL, V26, P1223
  • [2] Alfke K, 2004, AM J NEURORADIOL, V25, P584
  • [3] Endovascular occlusion of wide-necked aneurysms with a new intracranial microstent (neuroform) and detachable coils
    Benitez, RP
    Silva, MT
    Klem, J
    Veznedaroglu, E
    Rosenwasser, RH
    [J]. NEUROSURGERY, 2004, 54 (06) : 1359 - 1367
  • [4] Benndorf G, 2001, J ENDOVASC THER, V8, P566, DOI 10.1583/1545-1550(2001)008<0566:OSFTOA>2.0.CO
  • [5] 2
  • [6] Broadbent LP, 2003, AM J NEURORADIOL, V24, P1819
  • [7] A novel flexible endovascular stent for use in small and tortuous vessels
    Byrne, JV
    Bashiri, M
    Pasco, A
    Morris, JH
    [J]. NEURORADIOLOGY, 2000, 42 (01) : 56 - 61
  • [8] Intracranial stent placement to trap an extruded coil during endovascular aneurysm treatment: Technical note
    Fessler, RD
    Ringer, AJ
    Qureshi, AI
    Guterman, LR
    Hopkins, LN
    [J]. NEUROSURGERY, 2000, 46 (01) : 248 - 251
  • [9] Usefulness of the neuroform stent for the treatment of cerebral aneurysms: Results at initial (3-6-mo) follow-up
    Fiorella, D
    Albuquerque, FC
    Deshmukh, VR
    McDougall, CG
    [J]. NEUROSURGERY, 2005, 56 (06) : 1191 - 1201
  • [10] Fiorella D, 2004, AM J NEURORADIOL, V25, P1764