Variation in the levels of volatile trace gases within three hospital environments: implications for clinical breath testing

被引:22
作者
Boshier, P. R. [1 ]
Cushnir, J. R. [1 ]
Priest, O. H. [1 ]
Marczin, N. [2 ]
Hanna, G. B. [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, St Marys Hosp, Dept Surg & Canc, London W2 1NY, England
[2] Univ London Imperial Coll Sci Technol & Med, Dept Anaesthet Pain Med & Intens care, Chelsea & Westminster Hosp, London W2 1NY, England
关键词
TUBE MASS-SPECTROMETRY; EXHALED BREATH; SIFT-MS; ORGANIC-COMPOUNDS; PROPANOL;
D O I
10.1088/1752-7155/4/3/031001
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Selected ion flow tube mass spectrometry, SIFT-MS, has been used to determine the levels of volatile trace gases in the room air of three clinical environments within a busy teaching hospital. The main aims of this study were to establish background levels of trace gases and to compare them to levels typically found within exhaled breath. Over a period of one month, room air samples were collected daily from an outpatient clinic, a hospital ward and an operating theatre, in either the morning or the evening. The concentrations of seven volatile compounds were compared between different locations and different times of the day. Having established the baseline levels for these compounds, breath was collected from healthy volunteers working within each of the three clinical locations. The concentrations of isoprene, acetone, hydrogen cyanide and ammonia within room air samples collected from at least one of the three clinical environments were less than 25% of levels typically found in human breath. Based on the recommendation of previous authors these compounds may therefore be suitable for use as exhaled markers of disease. In comparison high levels of ethanol, propanol and acetic acid within room air samples collected from each location may diminish confidence in their use as breath biomarkers.
引用
收藏
页数:6
相关论文
共 16 条
[1]  
Amann A., 2005, Breath analysis for clinical diagnosis and therapeutic monitoring
[2]   Volatile organic compounds in the exhaled breath of young patients with cystic fibrosis [J].
Barker, M ;
Hengst, M ;
Schmid, J ;
Buers, HJ ;
Mittermaier, B ;
Klemp, D ;
Koppman, R .
EUROPEAN RESPIRATORY JOURNAL, 2006, 27 (05) :929-936
[3]   On the use of Tedlar® bags for breath-gas sampling and analysis [J].
Beauchamp, Jonathan ;
Herbig, Jens ;
Gutmann, Rene ;
Hansel, Armin .
JOURNAL OF BREATH RESEARCH, 2008, 2 (04)
[4]   Use of poly(ethylene terephtalate) film bag to sample and remove humidity from atmosphere containing volatile organic compounds [J].
Beghi, Sandra ;
Guillot, Jean-Michel .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1183 (1-2) :1-5
[5]   Repeatability of the Measurement of Exhaled Volatile Metabolites Using Selected Ion Flow Tube Mass Spectrometry [J].
Boshier, Piers R. ;
Marczin, Nandor ;
Hanna, George B. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2010, 21 (06) :1070-1074
[6]   Effects of ventilation on the collection of exhaled breath in humans [J].
Cope, KA ;
Watson, MT ;
Foster, WM ;
Sehnert, SS ;
Risby, TH .
JOURNAL OF APPLIED PHYSIOLOGY, 2004, 96 (04) :1371-1379
[7]   Nitric oxide from the human respiratory tract efficiently quantified by standardized single breath measurements [J].
Hogman, M ;
Stromberg, S ;
Schedin, U ;
Frostell, C ;
Hedenstierna, G ;
Gustafsson, LE .
ACTA PHYSIOLOGICA SCANDINAVICA, 1997, 159 (04) :345-346
[8]   Analysis of the isobaric compounds propanol, acetic acid and methyl formate in humid air and breath by selected ion flow tube mass spectrometry, SIFT-MS [J].
Pysanenko, Andriy ;
Spanel, Patrik ;
Smith, David .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2009, 285 (1-2) :42-48
[9]   Current status of clinical breath analysis [J].
Risby, T. H. ;
Solga, S. F. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2006, 85 (2-3) :421-426
[10]   Clinical application of breath biomarkers of oxidative stress status [J].
Risby, TH ;
Sehnert, SS .
FREE RADICAL BIOLOGY AND MEDICINE, 1999, 27 (11-12) :1182-1192