摘要
A relatively simple setup for collection and detection of light emitted from isolated photo-excited molecular ions has been constructed. It benefits from a high collection efficiency of photons, which is accomplished by using a cylindrical ion trap where one end-cap electrode is a mesh grid combined with an aspheric condenser lens. The geometry permits nearly 10% of the emitted light to be collected and, after transmission losses, approximately 5% to be delivered to the entrance of a grating spectrometer equipped with a detector array. The high collection efficiency enables the use of pulsed tunable lasers with low repetition rates (e.g., 20 Hz) instead of continuous wave (cw) lasers or very high repetition rate (e.g., MHz) lasers that are typically used as light sources for gas-phase fluorescence experiments on molecular ions. A hole has been drilled in the cylinder electrode so that a light pulse can interact with the ion cloud in the center of the trap. Simulations indicate that these modifications to the trap do not significantly affect the storage capability and the overall shape of the ion cloud. The overlap between the ion cloud and the laser light is basically 100%, and experimentally >50% of negatively charged chromophore ions are routinely photodepleted. The performance of the setup is illustrated based on fluorescence spectra of several laser dyes, and the quality of these spectra is comparable to those reported by other groups. Finally, by replacing the optical system with a channeltron detector, we demonstrate that the setup can also be used for gas-phase action spectroscopy where either depletion or fragmentation is monitored to provide an indirect measurement on the absorption spectrum of the ion. Published by AIP Publishing.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Negative gas-phase ion chemistry of silane: a quadrupole ion trap study
    Operti, Lorenza
    Rabezzana, Roberto
    Vaglio, Gian Angelo
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (18) : 2696 - 2700
  • [2] Enantioselective gas-phase ion-molecule reactions in a quadrupole ion trap
    Grigorean, G
    Gronert, S
    Lebrilla, CB
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 219 (01) : 79 - 87
  • [3] Gas-phase ion/ion interactions between peptides or proteins and iron ions in a quadrupole ion trap
    Payne, AH
    Glish, GL
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2001, 204 (1-3) : 47 - 54
  • [4] Empirical Calibration of a Cylindrical Ion Trap for Mass-Selected Gas-Phase Fluorescence Spectroscopy
    Lindkvist, Thomas Toft
    Langeland, Jeppe
    Kjaer, Christina
    Nielsen, Steen Brondsted
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2023, 35 (01) : 123 - 130
  • [5] Negative gas-phase ion chemistry of GeH4:: a quadrupole ion trap study
    Operti, L
    Rabezzana, R
    Turco, F
    Vaglio, GA
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (14) : 1963 - 1969
  • [6] A quadrupole ion trap mass spectrometer with three independent ion sources for the study of gas-phase ion/ion reactions
    Badman, Ethan R.
    Chrisman, Paul A.
    McLuckey, Scott A.
    Analytical Chemistry, 2002, 74 (24) : 6237 - 6243
  • [7] Nile blue shows its true colors in gas-phase absorption and luminescence ion spectroscopy
    Stockett, M. H.
    Houmoller, J.
    Nielsen, S. Brondsted
    JOURNAL OF CHEMICAL PHYSICS, 2016, 145 (10):
  • [8] Ion-molecule reactions in a quadrupole ion trap as a probe of the gas-phase structure of metal complexes
    Vachet, RW
    Hartman, JAR
    Callahan, JH
    JOURNAL OF MASS SPECTROMETRY, 1998, 33 (12): : 1209 - 1225
  • [9] Solvation in gas-phase reactions of sulfonic groups containing ionic liquids in electrospray ionization quadrupole ion trap mass spectrometry
    Cao, Xiaoji
    Yu, Yi
    Ye, Xueming
    Mo, Weiming
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2009, 15 (03) : 409 - 413
  • [10] A mass spectrometry and optical spectroscopy investigation of gas-phase ion formation in electrospray
    Chillier, XFD
    Monnier, A
    Bill, H
    Gulacar, FO
    Buchs, A
    McLuckey, SA
    Van Berkel, GJ
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1996, 10 (03) : 299 - 304