Failure analysis of fractured dental zirconia implants

被引:79
作者
Gahlert, M. [1 ]
Burtscher, D. [2 ]
Grunert, I. [2 ]
Kniha, H. [1 ]
Steinhauser, E. [3 ]
机构
[1] Private Dent Clin, D-80333 Munich, Germany
[2] Innsbruck Med Univ, MZA, Clin Dept Prosthet Dent, Innsbruck, Austria
[3] Munich Univ Appl Sci, Dept Precis & Microengn Engn Phys, Munich, Germany
关键词
failure analysis; fracture; mechanical failure; overloading occlusion; zirconia implant; COMMERCIALLY PURE TITANIUM; TISSUE; SURVIVAL; STRENGTH;
D O I
10.1111/j.1600-0501.2011.02206.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The purpose of the present study was the macroscopic and microscopic failure analysis of fractured zirconia dental implants. Methods: Thirteen fractured one-piece zirconia implants (Z-Look3) out of 170 inserted implants with an average in situ period of 36.75 +/- 5.34 months (range from 20 to 56 months, median 38 months) were prepared for macroscopic and microscopic (scanning electron microscopy [SEM]) failure analysis. These 170 implants were inserted in 79 patients. The patient histories were compared with fracture incidences to identify the reasons for the failure of the implants. Results: Twelve of these fractured implants had a diameter of 3.25 mm and one implant had a diameter of 4 mm. All fractured implants were located in the anterior side of the maxilla and mandibula. The patient with the fracture of the 4 mm diameter implant was adversely affected by strong bruxism. By failure analysis (SEM), it could be demonstrated that in all cases, mechanical overloading caused the fracture of the implants. Inhomogeneities and internal defects of the ceramic material could be excluded, but notches and scratches due to sandblasting of the surface led to local stress concentrations that led to the mentioned mechanical overloading by bending loads. Conclusions: The present study identified a fracture rate of nearly 10% within a follow-up period of 36.75 months after prosthetic loading. Ninety-two per cent of the fractured implants were so-called diameter reduced implants (diameter 3.25 mm). These diameter reduced implants cannot be recommended for further clinical use. Improvement of the ceramic material and modification of the implant geometry has to be carried out to reduce the failure rate of small-sized ceramic implants. Nevertheless, due to the lack of appropriate laboratory testing, only clinical studies will demonstrate clearly whether and how far the failure rate can be reduced.
引用
收藏
页码:287 / 293
页数:7
相关论文
共 40 条
  • [1] ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
  • [2] Fracture Strength of Zirconia Implants after Artificial Aging
    Andreiotelli, Marina
    Kohal, Ralf-Joachim
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2009, 11 (02) : 158 - 166
  • [3] Are ceramic implants a viable alternative to titanium implants? A systematic literature review
    Andreiotelli, Marina
    Wenz, Hans J.
    Kohal, Ralf-Joachim
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 : 32 - 47
  • [4] [Anonymous], 2009, 8436 DIN EN
  • [5] [Anonymous], 2000, 60672 DIN EN
  • [6] [Anonymous], 2007, International Standard ISO 14801-dentistry/implants-dynamic fatigue test for endosseous dental implants
  • [7] Balshi T J, 1996, Int J Oral Maxillofac Implants, V11, P660
  • [8] Binon PP, 2000, INT J ORAL MAX IMPL, V15, P76
  • [9] Burtscher D., 2009, 33 ANN C EUR PROSTH, V1-3, P48
  • [10] Treatment outcome of immediate, early and conventional single-tooth implants in the aesthetic zone: a systematic review to survival, bone level, soft-tissue, aesthetics and patient satisfaction
    den Hartog, Laurens
    Huddleston Slater, James J. R.
    Vissink, Arjan
    Meijer, Henny J. A.
    Raghoebar, Gerry M.
    [J]. JOURNAL OF CLINICAL PERIODONTOLOGY, 2008, 35 (12) : 1073 - 1086