A comparative surface topographical analysis of explanted total knee replacement prostheses: Oxidised zirconium vs cobalt chromium femoral components

被引:9
作者
Kennard, Emma [1 ]
Scholes, Susan C. [1 ]
Sidaginamale, Raghavendra [1 ]
Gangadharan, Rajkumar [2 ]
Weir, David J. [2 ]
Holland, James [2 ]
Deehan, David [2 ]
Joyce, Thomas J. [1 ]
机构
[1] Newcastle Univ, Sch Engn, Bioengn Grp, Stephenson Bldg, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
[2] Freeman Rd Hosp, Newcastle Upon Tyne NE1 7DN, Tyne & Wear, England
基金
英国工程与自然科学研究理事会;
关键词
Total knee replacement; Retrieval; Oxidised zirconium (OxZr); Surface roughness; Profilometry; RETRIEVAL ANALYSIS; POLYETHYLENE WEAR; FOLLOW-UP; ARTHROPLASTY; TKA; SIMULATOR; MINIMUM; ALLOY;
D O I
10.1016/j.medengphy.2017.10.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It has been proposed that an increased surface roughness of the femoral components of Total Knee Replacements (TKRs) may be a contributing factor to the accelerated wear of the polyethylene (PE) bearing and ultimately prosthesis failure. Oxidised Zirconium was introduced to the orthopaedic market in an attempt to reduce PE wear associated failures and increase the longevity of the prosthesis. In this study, non-contacting profilometry was used to measure the surface roughness of the femoral components of 6 retrieved TKRs (3 Oxidised Zirconium (OxZr) and 3 Cobalt Chromium alloy (CoCr) femoral components) and 2 as-manufactured femoral components (1 OxZr and 1 CoCr). A semi quantitative method was used to analyse the damage on the retrieved PE components. The S-a values for the retrieved OxZr femoral components (S-a = 0.093 mu m +/- 0.014) and for the retrieved CoCr femoral components (S-a = 0.065 mu m +/- 0.005) were significantly greater (p < .05) than the roughness values for the as-manufactured femoral components (OxZr S-a = 0.061 mu m +/- 0.004 and CoCr S-a = 0.042 mu m +/- 0.003). No significant difference was seen between the surface roughness parameters of the retrieved OxZr and retrieved CoCr femoral components. There was no difference between the PE component damage scores for the retrieved OxZr TKRs compared to the retrieved CoCr TKRs. These results agree with other studies that both OxZr and CoCr femoral components roughen during time in vivo but the lack of difference between the surface roughness measurements of the two materials is in contrast to previous topographical reports. Further analysis of retrieved OxZr TKRs is recommended so that a fuller appreciation of their benefits and limitations be obtained. (C) 2017 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 32 条
[1]  
[Anonymous], 13 NAT JOINT REG ENG
[2]  
[Anonymous], 2015, ANN REPORT
[3]  
[Anonymous], 1992, TRIBOLOGY FRICTION W
[4]  
[Anonymous], 2009, GEOM PROD SPEC GPS S
[5]  
Australian Orthopaedic Association National Joint Replacement Registry, 2016, ANN REPORT
[6]   Twenty-Year Survival Analysis in Total Knee Arthroplasty by a Single Surgeon [J].
Bae, Dae Kyung ;
Song, Sang Jun ;
Park, Man Jun ;
Eoh, Jae Hyung ;
Song, Jong Hoon ;
Park, Cheol Hee .
JOURNAL OF ARTHROPLASTY, 2012, 27 (07) :1297-1304
[7]   Performance assessment of femoral knee components made from cobalt-chromium alloy and oxidized zirconium [J].
Brandt, J. -M. ;
Guenther, L. ;
O'Brien, S. ;
Vecherya, A. ;
Turgeon, T. R. ;
Bohm, E. R. .
KNEE, 2013, 20 (06) :388-396
[8]   UHMWPE wear against roughened oxidized zirconium and CoCr femoral knee components during force-controlled simulation [J].
DesJardins, John D. ;
Burnikel, Brian ;
LaBerge, Martine .
WEAR, 2008, 264 (3-4) :245-256
[9]   Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator [J].
Ezzet, KA ;
Hermida, JC ;
Colwell, CW ;
D'Lima, DD .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :120-124
[10]   Osteolysis around total knee arthroplasty: A review of pathogenetic mechanisms [J].
Gallo, J. ;
Goodman, S. B. ;
Konttinen, Y. T. ;
Wimmer, M. A. ;
Holinka, M. .
ACTA BIOMATERIALIA, 2013, 9 (09) :8046-8058