The limits of determination for anions in water with the capillary electrophoresis method

被引:0
|
作者
Klojzy-Karczmarczyk, B [1 ]
Kmiecik, E [1 ]
Knap, W [1 ]
机构
[1] Stanislaw Staszic Univ Min & Met, Hydrol & Water Protect Dept, PL-30059 Krakow, Poland
来源
CHEMIA ANALITYCZNA | 2001年 / 46卷 / 03期
关键词
quality control; limits of decision; detection; determination; precision of results; capillary electrophoresis;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the case of capillary electrophoresis the limits of determination xo cannot be defined according to the accepted definition. The paper presents the approach applied and recommended by EURACHEM, relying on the approximation of the graph representing the relationships between the relative standard deviation RSD (%) and the concentration of given indicator in the spiked sample. Consistent with this approach the limit of determination is the lowest concentration of analyte in the sample which can be exactly quantified with precision, determined by the level of RSD (%).
引用
收藏
页码:377 / 385
页数:9
相关论文
共 50 条
  • [21] Determination of inorganic anions in Kraft pulping liquors by capillary electrophoresis
    Masselter, SM
    Zemann, AJ
    Bonn, GK
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1996, 19 (03): : 131 - 136
  • [22] Capillary electrophoresis for the determination of forensic anions in adulterated foods and beverages
    Soga, T
    Tajima, I
    Heiger, DN
    AMERICAN LABORATORY, 2000, 32 (03) : 124 - +
  • [23] Routine determination of major anions in atmospheric aerosols by capillary electrophoresis
    DabekZlotorzynska, E
    Piechowski, M
    Liu, F
    Kennedy, S
    Dlouhy, JF
    JOURNAL OF CHROMATOGRAPHY A, 1997, 770 (1-2) : 349 - 359
  • [24] Determination of inorganic and organic anions in beer and wort by capillary electrophoresis
    Soga, T
    Wakaura, M
    JOURNAL OF THE AMERICAN SOCIETY OF BREWING CHEMISTS, 1997, 55 (02) : 44 - 46
  • [25] Electroosmotic flow reversal for the determination of inorganic anions by capillary electrophoresis with methanol-water buffers
    Diress, AG
    Lucy, CA
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1027 (1-2) : 185 - 191
  • [26] Simultaneous determination of inorganic anions and cations in waters by capillary electrophoresis
    Padarauskas, A
    Olsauskaité, V
    Paliulionyté, V
    JOURNAL OF CHROMATOGRAPHY A, 1998, 829 (1-2) : 359 - 365
  • [27] Simultaneous determination of inorganic anions and organic acids by capillary electrophoresis
    Öztekin, N
    Erim, FB
    TURKISH JOURNAL OF CHEMISTRY, 2001, 25 (02) : 145 - 150
  • [28] Simultaneous Determination of Inorganic Anions and Cations in Water and Biological Samples by Capillary Electrophoresis with a Capacitive Coupled Contactless Conductivity Detector Using Capillary Filling Method
    Yamamoto, Sachio
    Fujiwara, Haruka
    Maruyama, Katsushi
    Tanaka, Yoshihide
    Kinoshita, Mitsuhiro
    Suzuki, Shigeo
    ANALYTICAL SCIENCES, 2019, 35 (03) : 295 - 300
  • [29] Simultaneous Determination of Inorganic Anions and Cations in Water and Biological Samples by Capillary Electrophoresis with a Capacitive Coupled Contactless Conductivity Detector Using Capillary Filling Method
    Sachio Yamamoto
    Haruka Fujiwara
    Katsushi Maruyama
    Yoshihide Tanaka
    Mitsuhiro Kinoshita
    Shigeo Suzuki
    Analytical Sciences, 2019, 35 : 295 - 300
  • [30] Capillary electrophoresis of inorganic anions
    Kaniansky, D
    Masár, M
    Marák, J
    Bodor, R
    JOURNAL OF CHROMATOGRAPHY A, 1999, 834 (1-2) : 133 - 178