Low dose scanning transmission electron microscopy of organic crystals by scanning moire fringes

被引:20
作者
S'ari, Mark [1 ]
Cattle, James [1 ]
Hondow, Nicole [1 ]
Brydson, Rik [1 ]
Brown, Andy [1 ]
机构
[1] Univ Leeds, Sch Chem & Proc Engn, Leeds LS2 9JT, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Low dose; Bright field STEM; Organic crystals; Scanning moire fringes; Dose-limited resolution; RADIATION-DAMAGE; SOLID DISPERSIONS; QUANTITATIVE-ANALYSIS; IMAGE-CONTRAST; BEAM DAMAGE; MICROGRAPHS; TEMPERATURE; STRAIN; STEM; DIFFRACTION;
D O I
10.1016/j.micron.2019.01.014
中图分类号
TH742 [显微镜];
学科分类号
摘要
In the pharmaceutical industry, it is important to determine the effects of crystallisation and processes, such as milling, on the generation of crystalline defects in formulated products. Conventional transmission electron microscopy and scanning transmission electron microscopy (STEM) can be used to obtain information on length scales unobtainable by other techniques, however, organic crystals are extremely susceptible to electron beam damage. This work demonstrates a bright field (BF) STEM method that can increase the information content per unit specimen damage by the use of scanning moire fringes (SMFs). SMF imaging essentially provides a magnification of the crystal lattice through the interference between closely aligned lattice fringes and a scanning lattice of similar spacing. The generation of SMFs is shown for three different organic crystals with varying electron beam sensitivity, theophylline, furosemide and felodipine. The electron fluence used to acquire the BF-STEM for the most sensitive material, felodipine was approximately 3.5 e(-)/angstrom(2). After one additional scan of felodipine (total fluence of approximately 7.0 e(-)/angstrom(2)), the SMFs were no longer visible due to extensive damage caused to the crystal. Irregularity in the SMFs suggested the presence of defects in all the organic crystals. Further effort is required to improve the data analysis and interpretation of the resulting SMF images, allowing more information regarding the crystal structure and defects to be extracted.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 54 条
[1]   Sub-angstrom resolution using aberration corrected electron optics [J].
Batson, PE ;
Dellby, N ;
Krivanek, OL .
NATURE, 2002, 418 (6898) :617-620
[2]   Surface area normalized dissolution to study differences in itraconazole-copovidone solid dispersions prepared by spray-drying and hot melt extrusion [J].
Bhardwaj, Vivekanand ;
Trasi, Niraj S. ;
Zemlyanov, Dmitry Y. ;
Taylor, Lynne S. .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2018, 540 (1-2) :106-119
[3]   High-resolution low-dose scanning transmission electron microscopy [J].
Buban, James P. ;
Ramasse, Quentin ;
Gipson, Bryant ;
Browning, Nigel D. ;
Stahlberg, Henning .
JOURNAL OF ELECTRON MICROSCOPY, 2010, 59 (02) :103-112
[4]  
BURSILL LA, 1981, NATURE, V292, P518, DOI 10.1038/292518a0
[5]   Studies on the crystallinity of a pharmaceutical development drug substance [J].
Byard, SJ ;
Jackson, SL ;
Smail, A ;
Bauer, M ;
Apperley, DC .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2005, 94 (06) :1321-1335
[6]   Robust uncertainty principles:: Exact signal reconstruction from highly incomplete frequency information [J].
Candès, EJ ;
Romberg, J ;
Tao, T .
IEEE TRANSACTIONS ON INFORMATION THEORY, 2006, 52 (02) :489-509
[7]   Transmission electron microscopy of a model crystalline organic, theophylline [J].
Cattle, J. ;
S'ari, M. ;
Hondow, N. ;
Abellan, P. ;
Brown, A. P. ;
Brydson, R. M. D. .
ELECTRON MICROSCOPY AND ANALYSIS GROUP CONFERENCE (EMAG2015), 2015, 644
[8]  
Cattle J, 2016, EUR MICR C 2016 P, P752, DOI [10.1002/9783527808465.EMC2016.6306, DOI 10.1002/9783527808465.EMC2016.6306]
[9]   Compressed sensing [J].
Donoho, DL .
IEEE TRANSACTIONS ON INFORMATION THEORY, 2006, 52 (04) :1289-1306
[10]   Transmission Electron Microscopy of Pharmaceutical Materials [J].
Eddleston, Mark D. ;
Bithell, Erica G. ;
Jones, William .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2010, 99 (09) :4072-4083