Real-Time Optical Fiber Dosimeter Probe

被引:4
作者
Croteau, Andre [1 ]
Caron, Serge [1 ]
Rink, Alexandra [2 ]
Jaffray, David [2 ]
Mermut, Ozzy [1 ]
机构
[1] INO, 2740 Einstein St, Quebec City, PQ G1P 4S4, Canada
[2] Univ Toronto, Princess Margaret Hosp, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
来源
OPTICAL FIBERS, SENSORS, AND DEVICES FOR BIOMEDICAL DIAGNOSTICS AND TREATMENT XI | 2011年 / 7894卷
关键词
Dosimetry; dosimeter probe; radiochromic thin films; real-time; optical fiber; passive; RADIOCHROMIC MEDIUM;
D O I
10.1117/12.873619
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is a pressing need for a passive optical fiber dosimeter probe for use in real-time monitoring of radiation dose delivered to clinical radiation therapy patients. An optical fiber probe using radiochromic material has been designed and fabricated based on a thin film of the radiochromic material on a dielectric mirror. Measurements of the net optical density vs. time before, during, and after irradiation at a rate of 500cGy/minute to a total dose of 5 Gy were performed. Net optical densities increased from 0.2 to 2.0 for radiochromic thin film thicknesses of 2 to 20 mu m, respectively.
引用
收藏
页数:7
相关论文
共 9 条
[1]   Investigation of the use of MOSFET for clinical IMRT dosimetric verification [J].
Chuang, CF ;
Verhey, LJ ;
Xia, P .
MEDICAL PHYSICS, 2002, 29 (06) :1109-1115
[2]  
Kron Tomas, 1994, Australasian Physical and Engineering Sciences in Medicine, V17, P175
[3]   Comparison of p-type commercial electron diodes for in vivo dosimetry [J].
Marre, D ;
Marinello, G .
MEDICAL PHYSICS, 2004, 31 (01) :50-56
[4]   Characterization and real-time optical measurements of the ionizing radiation dose response for a new radiochromic medium [J].
Rink, A ;
Vitkin, IA ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (08) :2510-2516
[5]   Suitability of radiochromic medium for real-time optical measurements of ionizing radiation dose [J].
Rink, A ;
Vitkin, IA ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (04) :1140-1155
[6]  
RINK A, 2008, Patent No. 7399977
[7]  
SCHULZ R J, 1983, Medical Physics (Woodbury), V10, P741, DOI 10.1118/1.595446
[8]   Response corrections for solid-state detectors in megavoltage photon dosimetry [J].
Yin, Z ;
Hugtenburg, RP ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (16) :3691-3702
[9]  
Yorke E., 2005, 87 AAPM TASK GROUP 6