Maximizing performance in supercritical fluid chromatography using low-density mobile phases

被引:14
作者
Gritti, Fabrice [1 ]
Fogwill, Michael [1 ]
Gilar, Martin [1 ]
Jarrell, Joseph A. [1 ]
机构
[1] Waters Corp, Instrument Core Res Fundamental, Milford, MA 01757 USA
关键词
Supercritical fluid chromatography; Low-density carbon dioxide; Column efficiency; Vacuum technology; Decoupling between inlet eluent; temperature and oven temperature; Insulating material; PRESSURE LIQUID-CHROMATOGRAPHY; THERMAL ENVIRONMENT; VISCOUS FRICTION; EFFICIENCY; COLUMNS; TEMPERATURE; DISSIPATION; SEPARATIONS; DISPERSION; MODELS;
D O I
10.1016/j.chroma.2016.09.024
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of a 3.0 mm x 150 mm column packed with 1.8 mu m fully porous HSS-SB-C-18 particles was investigated in supercritical fluid chromatography (SFC) with low-density, highly expansible carbon dioxide. These conditions are selected for the analysis of semi-volatile compounds. Elevated temperatures (>100 degrees C) were then combined with low column back pressures (<100 bar). In this work, the inlet temperature of pure carbon dioxide was set at 107 degrees C, the active back pressure regulator (ABPR) pressure was fixed at 100 bar, and the flow rate was set at 2.1 mL/min at 12 degrees C (liquefied carbon dioxide) and at an inlet column pressure close to 300 bar. Nine n-alkylbenzenes (from benzene to octadecylbenzene) were injected under linear (no sample overload) conditions. The severe steepness of the temperature gradients across the column diameter were predicted from a simplified heat transfer model. Such conditions dramatically lower the column performance by affecting the symmetry of the peak shape. In order to cope with this problem, three different approaches were experimentally tested. They include (1) the decoupling and the proper selection of the inlet eluent temperature with respect to the oven temperature, (2) the partial thermal insulation of the column using polyethylene aerogel, and (3) the application of a high vacuum (10(-5) Torr provided by a turbo-molecular pump) in a housing chamber surrounding the whole column body. The results reveal that (1) the column efficiency can be maximized by properly selecting the difference between the eluent and the oven temperatures, (2) the mere wrapping of the column with an excellent insulating material is insufficient to fully eliminate heat exchanges by conduction and the undesirable radial density gradients across the column i.d., and (3) the complete thermal insulation of the SFC column under high vacuum allows to maximize the column efficiency by maintaining the integrity of the peak shape. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:217 / 227
页数:11
相关论文
共 50 条
[21]   Enantioselective potential of polysaccharide-based chiral stationary phases in supercritical fluid chromatography [J].
Kucerova, Gabriela ;
Kalikova, Kveta ;
Tesarova, Eva .
CHIRALITY, 2017, 29 (06) :239-246
[22]   Characterization of stationary phases by a linear solvation energy relationship utilizing supercritical fluid chromatography [J].
Mitchell, Clifford R. ;
Benz, Nancy J. ;
Zhang, Shuhong .
JOURNAL OF SEPARATION SCIENCE, 2010, 33 (19) :3060-3067
[23]   Performance of the same column in supercritical fluid chromatography and in liquid chromatography [J].
Lambert, Nandor ;
Felinger, Attila .
JOURNAL OF CHROMATOGRAPHY A, 2015, 1409 :234-240
[24]   EFFICIENCY IN PACKED-COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY USING A MODIFIED MOBILE PHASE [J].
BERGER, TA ;
DEYE, JF .
CHROMATOGRAPHIA, 1991, 31 (11-12) :529-534
[25]   Comparison of ultra-high performance supercritical fluid chromatography and ultra-high performance liquid chromatography for the analysis of pharmaceutical compounds [J].
Perrenoud, Alexandre Grand-Guillaume ;
Veuthey, Jean-Luc ;
Guillarme, Davy .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1266 :158-167
[26]   Pharmaceutical-enantiomers resolution using immobilized polysaccharide-based chiral stationary phases in supercritical fluid chromatography [J].
De Klerck, Katrijn ;
Vander Heyden, Yvan ;
Mangelings, Debby .
JOURNAL OF CHROMATOGRAPHY A, 2014, 1328 :85-97
[27]   Effects of supercritical fluid chromatography conditions on enantioselectivity and performance of polyproline-derived chiral stationary phases [J].
Novell, Arnau ;
Mendez, Alberto ;
Minguillon, Cristina .
JOURNAL OF CHROMATOGRAPHY A, 2015, 1403 :138-143
[28]   Enantioseparation on AminoAlcohol-DerivedStationary Phases by Supercritical Fluid Chromatography and High-Performance Liquid Chromatograpy [J].
Kim, Gyuri ;
Ryoo, Jae Jeong .
BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2020, 41 (09) :943-947
[29]   SULFUR-DIOXIDE AS A MOBILE PHASE IN SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
LEREN, E ;
LANDMARK, KE ;
GREIBROKK, T .
CHROMATOGRAPHIA, 1991, 31 (11-12) :535-538
[30]   Low aspect ratio packed capillary columns in supercritical fluid chromatography [J].
Li, WB ;
Pyo, DJ ;
Wan, YJ ;
Ibanez, E ;
Malik, A ;
Lee, ML .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1996, 8 (04) :259-268