Testing physiologic monitor alarm customization software to reduce alarm rates and improve nurses' experience of alarms in a medical intensive care unit

被引:37
作者
Ruppel, Halley [1 ,5 ]
De Vaux, Laura [2 ]
Cooper, Dawn [2 ]
Kunz, Steffen [3 ]
Duller, Bernd [4 ]
Funk, Marjorie [1 ,2 ]
机构
[1] Yale Sch Nursing, Yale West Campus, West Haven, CT 06477 USA
[2] Yale New Haven Med Ctr, 20 York St, New Haven, CT 06504 USA
[3] Philips Med Syst Boblingen GmbH, Boblingen, Germany
[4] Perpet Prod, Stuttgart, Germany
[5] Kaiser Permanente Northen Calif, Div Res, Oakland, CA USA
关键词
NUISANCE ALARMS; FATIGUE;
D O I
10.1371/journal.pone.0205901
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background Clinicians in intensive care units experience alarm fatigue related to frequent false and non-actionable alarms produced by physiologic monitors. To reduce non-actionable alarms, alarm settings may need to be customized for individual patients; however, nurses may not customize alarms because of competing demands and alarm fatigue. Objective To examine the effectiveness and acceptance of physiologic monitor software to support customization of alarms. Methods This pre/post intervention study was conducted in a 56-bed medical intensive care unit. IntelliVue (R) Alarm Advisor customization support software for alarm limit violations was installed on all monitors and education on its use provided. For 2 months before and after implementation of the software, data were collected on patient characteristics from the electronic health record, alarm counts and duration from the monitoring system, and nurses' experience of alarms from a survey. Results Medium-priority heart rate, respiratory rate, and arterial pressure alarms were significantly reduced after software implementation (9.3%, 11.8%, and 15.9% reduction respectively; p<0.001 for all). The duration of these alarms was also significantly shorter (7.8%, 13.3%, and 9.3% reduction respectively; p<0.05 for all). The number and duration of SpO(2) alarms did not decrease (p>0.05 for both). Patients post-intervention had worse Glasgow Coma Scale scores (p = 0.014), but otherwise were comparable to those pre-intervention. Nurses reported less time spent on non-actionable alarms post-intervention than pre-intervention (p = 0.026). Also lower post-intervention were the proportions of nurses who reported that alarms disturbed their workflow (p = 0.027) and who encountered a situation where an important alarm was ignored (p = 0.043). The majority (>50%) agreed that the software supported setting appropriate alarm limits and was easy to use. Conclusion Alarm customization software was associated with a reduction in alarms. Use of software to support nurses' recognition of trends in patients' alarms and facilitate changes to alarm settings may add value to alarm reduction initiatives.
引用
收藏
页数:16
相关论文
共 23 条
[1]  
[Anonymous], 2013, Joint Commission Perspectives, V35, P1
[2]   Video Analysis of Factors Associated With Response Time to Physiologic Monitor Alarms in a Children's Hospital [J].
Bonafide, Christopher P. ;
Localio, Russell ;
Holmes, John H. ;
Nadkarni, Vinay M. ;
Stemler, Shannon ;
MacMurchy, Matthew ;
Zander, Miriam ;
Roberts, Kathryn E. ;
Lin, Richard ;
Keren, Ron .
JAMA PEDIATRICS, 2017, 171 (06) :524-531
[3]   Clinical Trial of an Educational Program to Decrease Monitor Alarms in a Medical Intensive Care Unit [J].
Brantley, Arian ;
Collins-Brown, Sandra ;
Kirkland, Jasmine ;
Knapp, Meghan ;
Pressley, Jackie ;
Higgins, Melinda ;
Mcmurtry, James P. .
AACN ADVANCED CRITICAL CARE, 2016, 27 (03) :283-289
[4]   Effect of altering alarm settings: A randomized controlled study [J].
Cvach, Maria ;
Rothwell, Kathleen J. ;
Cullen, Ann Marie ;
Nayden, Mary Grace ;
Cvach, Nicholas ;
Pham, Julius Cuong .
Biomedical Instrumentation and Technology, 2015, 49 (03) :214-222
[5]   Daily Electrode Change and Effect on Cardiac Monitor Alarms An Evidence-Based Practice Approach [J].
Cvach, Maria M. ;
Biggs, Madalyn ;
Rothwell, Kathleen J. ;
Charles-Hudson, Charmaine .
JOURNAL OF NURSING CARE QUALITY, 2013, 28 (03) :265-271
[6]   Reduction of nonactionable alarms in medical intensive care [J].
De Vaux L. ;
Cooper D. ;
Knudson K. ;
Gasperini M. ;
Rodgerson K. ;
Funk M. .
Biomedical Instrumentation and Technology, 2017, 51 (Horizons) :58-61
[7]   Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients [J].
Drew, Barbara J. ;
Harris, Patricia ;
Zegre-Hemsey, Jessica K. ;
Mammone, Tina ;
Schindler, Daniel ;
Salas-Boni, Rebeca ;
Bai, Yong ;
Tinoco, Adelita ;
Ding, Quan ;
Hu, Xiao .
PLOS ONE, 2014, 9 (10)
[8]   Understanding heart rate alarm adjustment in the intensive care units through an analytical approach [J].
Fidler, Richard L. ;
Pelter, Michele M. ;
Drew, Barbara J. ;
Palacios, Jorge Arroyo ;
Bai, Yong ;
Stannard, Daphne ;
Aldrich, J. Matt ;
Hu, Xiao .
PLOS ONE, 2017, 12 (11)
[9]   Safety Analysis of Proposed Data-Driven Physiologic Alarm Parameters for Hospitalized Children [J].
Goel, Veena V. ;
Poole, Sarah F. ;
Longhurst, Christopher A. ;
Platchek, Terry S. ;
Pageler, Natalie M. ;
Sharek, Paul J. ;
Palma, Jonathan P. .
JOURNAL OF HOSPITAL MEDICINE, 2016, 11 (12) :817-823
[10]   MONITOR ALARM FATIGUE: STANDARDIZING USE OF PHYSIOLOGICAL MONITORS AND DECREASING NUISANCE ALARMS [J].
Graham, Kelly Creighton ;
Cvach, Maria .
AMERICAN JOURNAL OF CRITICAL CARE, 2010, 19 (01) :28-34