Patient-specific glenoid guides provide accuracy and reproducibility in total shoulder arthroplasty

被引:85
作者
Gauci, M. O. [1 ,2 ]
Boileau, P. [1 ,2 ]
Baba, M. [1 ,3 ]
Chaoui, J. [1 ,4 ]
Walch, G. [1 ,5 ]
机构
[1] CHU Nice, Nice, France
[2] Inst Univ Locomoteur & Sport, Hop Pasteur 2,30 Voie Romaine, F-06000 Nice, France
[3] Specialty Orthopaed, Sydney, NSW, Australia
[4] Telecom Brittany, 655 Ave Technopole, F-29200 Plouzane, France
[5] Ctr Orthoped Santy, Unite Epaule, 24 Ave Paul Santy, F-69008 Lyon, France
关键词
COMPONENT PLACEMENT; VERSION; POSITION; SURGEON;
D O I
10.1302/0301-620X.98B8.37257
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aims Patient-specific glenoid guides ( PSGs) claim an improvement in accuracy and reproducibility of the positioning of components in total shoulder arthroplasty ( TSA). The results have not yet been confirmed in a prospective clinical trial. Our aim was to assess whether the use of PSGs in patients with osteoarthritis of the shoulder would allow accurate and reliable implantation of the glenoid component. Patients and Methods A total of 17 patients ( three men and 14 women) with a mean age of 71 years ( 53 to 81) awaiting TSA were enrolled in the study. Pre- and post-operative version and inclination of the glenoid were measured on CT scans, using 3D planning automatic software. During surgery, a congruent 3D-printed PSG was applied onto the glenoid surface, thus determining the entry point and orientation of the central guide wire used for reaming the glenoid and the introduction of the component. Manual segmentation was performed on post-operative CT scans to compare the planned and the actual position of the entry point ( mm) and orientation of the component (degrees). Results The mean error in the accuracy of the entry point was -0.1 mm ( standard deviation ( SD) 1.4) in the horizontal plane, and 0.8 mm ( SD 1.3) in the vertical plane. The mean error in the orientation of the glenoid component was 3.4 degrees ( SD 5.1 degrees) for version and 1.8 degrees ( SD 5.3 degrees) for inclination. Conclusion Pre-operative planning with automatic software and the use of PSGs provides accurate and reproducible positioning and orientation of the glenoid component in anatomical TSA.
引用
收藏
页码:1080 / 1085
页数:6
相关论文
共 21 条
  • [1] Variability of measurement of glenoid version on computed tomography scan
    Bokor, DJ
    O'Sullivan, MD
    Hazan, GJ
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 1999, 8 (06) : 595 - 598
  • [2] Two-Dimensional Glenoid Version Measurements Vary with Coronal and Sagittal Scapular Rotation
    Bryce, Chris D.
    Davison, Andrew C.
    Lewis, Gregory S.
    Wang, Li
    Flemming, Donald J.
    Armstrong, April D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (03) : 692 - 699
  • [3] Comparison of standard two-dimensional and three-dimensional corrected glenoid version measurements
    Budge, Matthew D.
    Lewis, Gregory S.
    Schaefer, Eric
    Coquia, Stephanie
    Flemming, Donald J.
    Armstrong, April D.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2011, 20 (04) : 577 - 583
  • [4] Chaoui J, 2010, CAOS S, P98
  • [5] Risks of loosening of a prosthetic glenoid implanted in retroversion
    Farron, Alain
    Terrier, Alexandre
    Buechler, Philippe
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2006, 15 (04) : 521 - 526
  • [6] Predicting normal glenoid version from the pathologic scapula: a comparison of 4 methods in 2- and 3-dimensional models
    Ganapathi, Asvin
    McCarron, Jesse A.
    Chen, Xi
    Iannotti, Joseph P.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2011, 20 (02) : 234 - 244
  • [7] Gazielly DF, 2007, MAITRISE ORTHOPEDIQU, V162, P6
  • [8] Comparison of Patient-Specific Instruments with Standard Surgical Instruments in Determining Glenoid Component Position A Randomized Prospective Clinical Trial
    Hendel, Michael D.
    Bryan, Jason A.
    Barsoum, Wael K.
    Rodriguez, Eric J.
    Brems, John J.
    Evans, Peter J.
    Iannotti, Joseph P.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (23) : 2167 - 2175
  • [9] Ho JC., 2013, J Bone Joint Surg Am, V95, P82
  • [10] Effect of glenoid deformity on glenoid component placement in primary shoulder arthroplasty
    Iannotti, Joseph P.
    Greeson, Clay
    Downing, Devin
    Sabesan, Vani
    Bryan, Jason A.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (01) : 48 - 55