A phantom study of the geometric accuracy of computed tomographic and magnetic resonance imaging stereotactic localization with the Leksell stereotactic system

被引:86
|
作者
Yu, C
Apuzzo, MLJ
Zee, CS
Petrovich, Z
机构
[1] Univ So Calif, Keck Sch Med, Dept Radiat Oncol, Los Angeles, CA 90033 USA
[2] Univ So Calif, Keck Sch Med, Dept Neurol Surg, Los Angeles, CA 90033 USA
[3] Univ So Calif, Keck Sch Med, Dept Radiol, Los Angeles, CA 90033 USA
关键词
geometric accuracy; image distortion; radiosurgery; stereotactic localization;
D O I
10.1097/00006123-200105000-00025
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To assess the spatial accuracy of magnetic resonance imaging (MRI) and computed tomographic stereotactic localization with the Leksell stereotactic system. METHODS: The phantom was constructed in the shape of a box, 164 mm in each dimension, with three perpendicular arrays of solid acrylic rod, 5 mm in diameter and spaced 30 mm apart within the phantom. In this study, images from two different MRI scanners and a computed tomographic scanner were obtained using the same Leksell (Elekta Instruments, Stockholm, Sweden) head frame placement. The coordinates of the rod images in the three principal planes were measured by using a tool provided with Leksell GammaPlan software (Elekta Instruments, Norcross, CA) and were compared with the physical phantom measurements. RESULTS: The greatest distortion was found around the periphery, and the least distortion (<1.5 mm) was present in the middle and most other areas of the phantom. In the phantom study using computed tomography, the mean values of the maximum errors for the x, y, and z axes were 1.0 mm (range, 0.2-1.3 mm), 0.4 mm (range, 0.1-0.8 mm), and 3.8 mm (range, 1.9-5.1 mm), respectively. The mean values of the maximum errors when using the Philips MRI scanner (Philips Medical Systems, Shelton, CT) were 0.9 mm (range, 0.4-1.7 mm), 0.2 mm (range, 0.0-0.7 mm), and 1.9 mm (range, 1.3-2.3 mm), respectively. Using the Siemens MRI scanner (Siemens Medical Systems, New York, NY), these values were 0.4 mm (range, 0.0-0.7 mm), 0.6 mm (range, 0.0-1.0 mm), and 1.6 mm (range, 0.8-2.0 mm), respectively. The geometric accuracy of the MRI scans when using the Siemens scanner was greatly improved after the implementation of a new software patch provided by the manufacturer. The accuracy also varied with the direction of phase encoding. CONCLUSION: The accuracy of target localization for most intracranial lesions during stereotactic radiosurgery can be achieved within the size of a voxel, especially by using the Siemens MRI scanner at current specifications and with a new software patch. However, caution is warranted when imaging peripheral lesions, where the distortion is greatest.
引用
收藏
页码:1092 / 1098
页数:7
相关论文
共 50 条
  • [2] A Phantom Study of the Spatial Precision and Accuracy of Stereotactic Localization Using Computed Tomography Imaging with the Leksell Stereotactic System
    Tafreshi, Ali R.
    Peng, Terrance
    Yu, Cheng
    Kramer, Daniel R.
    Gogia, Angad S.
    Lee, Morgan B.
    Barbaro, Michael F.
    Sebastian, Rinu
    Del Campo-Vera, Roberto Martin
    Chen, Kuang-Hsuan
    Kellis, Spencer S.
    Lee, Brian
    WORLD NEUROSURGERY, 2020, 139 : E297 - E307
  • [3] A phantom study to assess the accuracy of stereotactic localization, using T1-weighted magnetic resonance imaging with the Leksell stereotactic system
    Walton, L
    Hampshire, A
    Forster, DMC
    Kemeny, AA
    NEUROSURGERY, 1996, 38 (01) : 170 - 176
  • [4] An in vitro study to evaluate the accuracy of stereotactic localization using magnetic resonance imaging by means of the Leksell Stereotactic System
    Gliemroth, J
    Gaebel, C
    Kehler, U
    Grande-Nagel, I
    Missler, U
    Arnold, H
    MINIMALLY INVASIVE NEUROSURGERY, 2002, 45 (01) : 1 - 5
  • [6] Development of a unique phantom to assess the geometric accuracy of magnetic resonance imaging for stereotactic localization
    Orth, RC
    Sinha, P
    Madsen, EL
    Frank, G
    Korosec, FR
    Mackie, TR
    Mehta, MP
    NEUROSURGERY, 1999, 45 (06) : 1423 - 1429
  • [7] Development of a unique phantom to assess the geometric accuracy of magnetic resonance imaging for stereotactic localization - Comment
    Loeffler, JS
    Kooy, HM
    Chapman, P
    NEUROSURGERY, 1999, 45 (06) : 1430 - 1431
  • [8] The accuracy of magnetic resonance imaging for stereotactic localization
    Yu, C
    Apuzzo, MLJ
    Zee, CS
    Luxton, G
    Petrovich, Z
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2699 - 2702
  • [9] An algorithm for stereotactic localization by computed tomography or magnetic resonance imaging
    Dai, JR
    Zhu, YP
    Qu, HM
    Hu, YM
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01): : N1 - N7
  • [10] Stereotactic target localization accuracy in interventional magnetic resonance imaging
    Samset, E
    Hirschberg, H
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2002, 79 (3-4) : 191 - 201