Polyethylene wear in Scanhip(R) arthroplasty with a 22 or 32 mm head - 62 matched patients followed for 7-9 years

被引:38
作者
Kesteris, U
Ilchmann, T
Wingstrand, H
Onnerfalt, R
机构
[1] Department of Orthopedics, Lund University Hospital, 5-221 85, Lund
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1996年 / 67卷 / 02期
关键词
D O I
10.3109/17453679608994655
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We measured radiographic polyethylene wear in patients with Scanhip (R) arthroplasty and no clinical or radiographic signs of loosening. The patients were divided into 2 groups according to head sizes. 32 patients (33 hips) had an implant with a 22 mm and 30 patients (34 hips) with a 32 mm head. They were followed for 7-9 years. The groups were matched for diagnosis, sex, weight, age, and time of follow-up, The mean linear wear with a 22 mm head was 1.1 mm and with a 32 mm head 1.5 mm (p 0.004), which corresponds to a yearly wear rate of 0.15 mm and 0.18 mm, respectively. The mean difference in volumetric wear was greater, 420 mm(3), as compared to 1239 mm(3).
引用
收藏
页码:125 / 127
页数:3
相关论文
共 21 条
[1]  
[Anonymous], CLIN ORTHOPAEDICS
[2]  
Bankston A B, 1994, J Arthroplasty, V9, P533, DOI 10.1016/0883-5403(94)90101-5
[3]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]  
CALLAGHAN JJ, 1995, CLIN ORTHOP RELAT R, V317, P14
[5]  
CAMPBELL P, 1995, THESIS U GOTHENBURG
[6]  
CARLSSON AS, 1984, CLIN ORTHOP RELAT R, V185, P145
[7]   OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[8]  
Charnley J, 1973, CLIN ORTHOP RELAT R, V95, P9
[9]  
CLARKE IC, 1976, CLIN ORTHOP RELAT R, P126
[10]  
GRIFFITH MJ, 1978, CLIN ORTHOP RELAT R, V137, P37