High-resolution hard X-ray and gamma-ray spectrometers based on superconducting absorbers coupled to superconducting transition edge sensors

被引:13
作者
van den Berg, ML [1 ]
Chow, DT [1 ]
Loshak, A [1 ]
Cunningham, MF [1 ]
Barbee, TW [1 ]
Frank, M [1 ]
Labov, SE [1 ]
机构
[1] Univ Calif Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
来源
X-RAY AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY XI | 2000年 / 4140卷
关键词
microcalorimeter; superconducting absorbers; x-ray spectroscopy; gamma-ray spectroscopy;
D O I
10.1117/12.409140
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
We are developing detectors based on bulk superconducting absorbers coupled to superconducting transition edge sensors (TES) for high-resolution spectroscopy of hard X-rays and soft gamma-rays. We have achieved an energy resolution of 70 eV FWHM at 60 keV using a 1 x 1 x 0.25 mm(3) Sn absorber coupled to a Mo/Cu multilayer TES with a transition temperature of 100 mK. The response of this detector is compared with a simple model using only material properties data and characteristics derived from IV-measurements. We have also manufactured detectors using superconducting absorbers with a higher stopping power, such as Pb and Ta. We present our first measurements of these detectors, including the thermalization characteristics of the bulk superconducting absorbers. The differences in performance between the detectors are discussed and an outline of the future direction of our detector development efforts is given.
引用
收藏
页码:436 / 444
页数:9
相关论文
共 22 条
  • [1] High energy resolution bolometers for nuclear physics and x-ray spectroscopy
    Alessandrello, A
    Beeman, JW
    Brofferio, C
    Cremonesi, O
    Fiorini, E
    Giuliani, A
    Haller, EE
    Monfardini, A
    Nucciotti, A
    Pavan, M
    Pessina, G
    Previtali, E
    Zanotti, L
    [J]. PHYSICAL REVIEW LETTERS, 1999, 82 (03) : 513 - 515
  • [2] SELECTION OF A THERMAL BONDING AGENT FOR TEMPERATURES BELOW 1K
    ANDERSON, AC
    PETERSON, RE
    [J]. CRYOGENICS, 1970, 10 (05) : 430 - &
  • [3] Low-temperature X-ray detectors for precise Lamb shift measurements on hydrogen-like heavy ions
    Bleile, A
    Egelhof, P
    Kluge, HJ
    Liebisch, U
    McCammon, D
    Meier, HJ
    Sebastián, O
    Stahle, CK
    Weber, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) : 488 - 491
  • [4] QUASI-PARTICLE TRAPPING AND THE QUASI-PARTICLE MULTIPLIER
    BOOTH, NE
    [J]. APPLIED PHYSICS LETTERS, 1987, 50 (05) : 293 - 295
  • [5] Gamma-ray spectrometers using a bulk Sn absorber coupled to a Mo/Cu multilayer superconducting transition edge sensor
    Chow, DT
    Lindeman, MA
    Cunningham, MF
    Frank, M
    Barbee, TW
    Labov, SE
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) : 196 - 200
  • [6] CHOW DT, 2000, IN PRESS P SPIE, V4141
  • [7] FURTHER RESULTS ON MU-CALORIMETERS WITH SUPERCONDUCTING ABSORBER
    COSULICH, E
    GATTI, F
    VITALE, S
    [J]. JOURNAL OF LOW TEMPERATURE PHYSICS, 1993, 93 (3-4) : 263 - 268
  • [8] HYDROGEN CONTAMINATION IN TANTALUM AND NIOBIUM FOLLOWING UHV-DEGASSING
    FABER, K
    SCHULTZ, H
    [J]. SCRIPTA METALLURGICA, 1972, 6 (11): : 1065 - &
  • [9] CALORIMETERS WITH PROXIMITY-EFFECT THERMOMETERS AND LEAD AND TIN ABSORBERS
    FORSTER, G
    HALLATSCHEK, K
    WILHELM, D
    PROBST, F
    SEIDEL, W
    KELLNER, E
    VONFEILITZSCH, F
    [J]. JOURNAL OF LOW TEMPERATURE PHYSICS, 1993, 93 (3-4) : 219 - 224
  • [10] SUPERCONDUCTING RHENIUM AS ABSORBER FOR THERMAL DETECTORS
    GATTI, F
    VITALE, S
    BARABINO, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 315 (1-3) : 260 - 262