CdTe and CdZnTe room-temperature X-ray and gamma ray detectors and imaging systems

被引:13
|
作者
Eisen, Y [1 ]
Shor, A [1 ]
机构
[1] Soreq NRC, IL-81800 Yavne, Israel
关键词
D O I
10.1557/PROC-487-129
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
CdTe and CdZnTe X-ray and gamma ray detectors in the form of single elements or as monolithic segmented arrays have been shown to be useful in imaging systems utilized in medical, research or industrial applications. These detectors possess inherently better energy resolution than scintillators coupled to either photodiodes or photomultipliers, and they may lead to compact imaging systems or to imaging systems of enhanced spatial resolution and better contrast resolution. Photopeak efficiencies of these detectors is greatly affected by relatively low hole mobility-lifetime product and therefore continuing efforts are still underway to improve the characteristics of both CdTe and CdZnTe materials in order to achieve reproducible detectors with higher photopeak efficiencies for either low or high energy gamma rays. The following paper is divided into three parts: The first part compares the characteristics of planar CdTe and CdZnTe single elements nuclear detectors containing metal contacts. Characteristics include: Charge collection efficiencies for both electrons and holes indicated by the mobility-lifetime product, energy resolutions, leakage currents and robustness in field use. The second part describes excellent spectroscopic results using a 1cm thick CdZnTe monolithic segmented pad detector array. This part also compares spectra for various gamma energies obtained by this segmented detector to that of a 1cm(3) detector acting as a single element planar detector. The third part discusses the characteristics of a new generation nuclear gamma camera for medical diagnostics based on room-temperature CdTe and CdZnTe spectrometers and its advantages over an Anger type scintillating nuclear camera.
引用
收藏
页码:129 / 145
页数:17
相关论文
共 50 条
  • [1] CdTe and CdZnTe materials for room-temperature X-ray and gamma ray detectors
    Eisen, Y
    Shor, A
    JOURNAL OF CRYSTAL GROWTH, 1998, 184 : 1302 - 1312
  • [2] CdTe and CdZnTe X-ray and gamma-ray detectors for imaging systems
    Eisen, Y
    Shor, A
    Mardor, I
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) : 1191 - 1198
  • [3] CdTe and CdZnTe X-ray and gamma-ray detectors for imaging systems
    Eisen, Y
    Shor, A
    Mardor, I
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 3311 - 3319
  • [4] CdTe and CdZnTe crystals for room temperature gamma-ray detectors
    Franc, J
    Höschl, P
    Belas, E
    Grill, R
    Hlídek, P
    Moravec, P
    Bok, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 434 (01): : 146 - 151
  • [5] A System for the Characterization and Testing of CdZnTe/CdTe Pixel Detectors for X-ray and Gamma-ray Imaging
    Arodzero, Anatoli
    Barber, William C.
    Damron, Matthew Q.
    Hartsough, Neal E.
    Iwanczyk, Jan S.
    Malakhov, Nail
    Nygard, Einar
    Moraes, Danielle
    Weilhammer, Peter
    Jarron, Pierre
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3638 - 3642
  • [6] CdTe and CdZnTe gamma ray detectors for medical and industrial imaging systems
    Eisen, Y
    Shor, A
    Mardor, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 428 (01): : 158 - 170
  • [7] CdTe and CdZnTe gamma ray detectors for medical and industrial imaging systems
    Soreq NRC, 81800 Yavne, Israel
    Nucl Instrum Methods Phys Res Sect A, 1 (158-170):
  • [8] Post-growth thermal annealing study of CdZnTe for developing room-temperature X-ray and gamma-ray detectors
    Yang, G.
    Bolotnikov, A. E.
    Fochuk, P. M.
    Kopach, O.
    Franc, J.
    Belas, E.
    Kim, K. H.
    Camarda, G. S.
    Hossain, A.
    Cui, Y.
    Adams, A. L.
    Radja, A.
    Pinder, R.
    James, R. B.
    JOURNAL OF CRYSTAL GROWTH, 2013, 379 : 16 - 20
  • [9] Dynamics of native oxide growth on CdTe and CdZnTe X-ray and gamma-ray detectors
    Zazvorka, Jakub
    Franc, Jan
    Beran, Lukas
    Moravec, Pavel
    Pekarek, Jakub
    Veis, Martin
    SCIENCE AND TECHNOLOGY OF ADVANCED MATERIALS, 2016, 17 (01) : 792 - 798
  • [10] Characterization of CdTe and CdZnTe detectors for gamma-ray imaging applications
    Verger, L
    Boitel, M
    Gentet, MC
    Hamelin, R
    Mestais, C
    Mongellaz, F
    Rustique, J
    Sanchez, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2): : 297 - 309