Noise considerations of three-point water-fat separation imaging methods

被引:7
作者
Wen, Zhifei [1 ,2 ]
Reeder, Scott B. [2 ]
Pineda, Angel R. [2 ]
Pelc, Norbert J. [2 ,3 ]
机构
[1] Stanford Univ, Dept Phys, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
关键词
MRI; water-fat separation; three point; NSA; CRB; simulation;
D O I
10.1118/1.2952644
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Separation of water from fat tissues in magnetic resonance imaging is important for many applications because signals from fat tissues often interfere with diagnoses that are usually based on water signal characteristics. Water and fat can be separated with images acquired at different echo time shifts. The three-point method solves for the unknown off-resonance frequency together with the water and fat densities. Noise performance of the method, quantified by the effective number of signals averaged (NSA), is an important metric of the water and fat images. The authors use error propagation theory and Monte Carlo simulation to investigate two common reconstructive approaches: an analytic-solution based estimation and a least-squares estimation. Two water-fat chemical shift (CS) encoding strategies, the symmetric (-theta,0,theta) and the shifted (0,theta,2 theta) schemes are studied and compared. Results show that NSAs of water and fat can be different and they are dependent on the ratio of intensities of the two species and each of the echo time shifts. The NSA is particularly poor for the symmetric (-theta,0,theta) CS encoding when the water and fat signals are comparable. This anomaly with equal amounts of water and fat is analyzed in a more intuitive geometric illustration. Theoretical prediction of NSA matches well with simulation results at high signal-to-noise ratio (SNR), while deviation arises at low SNR, which suggests that Monte Carlo simulation may be more appropriate to accurately predict noise performance of the algorithm when SNR is low. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:3597 / 3606
页数:10
相关论文
共 16 条
  • [1] Understanding phase maps in MRI: A new cutline phase unwrapping method
    Chavez, S
    Xiang, QS
    An, L
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (08) : 966 - 977
  • [2] An automated iterative algorithm for water and fat decomposition in three-point Dixon magnetic resonance imaging
    Chen, QS
    Schneider, E
    Aghazadeh, B
    Weinhous, MS
    Humm, J
    Ballon, D
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2341 - 2347
  • [3] SIMPLE PROTON SPECTROSCOPIC IMAGING
    DIXON, WT
    [J]. RADIOLOGY, 1984, 153 (01) : 189 - 194
  • [4] MULTIPOINT DIXON TECHNIQUE FOR WATER AND FAT PROTON AND SUSCEPTIBILITY IMAGING
    GLOVER, GH
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (05): : 521 - 530
  • [5] 3-POINT DIXON TECHNIQUE FOR TRUE WATER FAT DECOMPOSITION WITH BO INHOMOGENEITY CORRECTION
    GLOVER, GH
    SCHNEIDER, E
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) : 371 - 383
  • [6] Method for efficient fast spin echo Dixon imaging
    Ma, JF
    Singh, SK
    Kumar, AJ
    Leeds, NE
    Broemeling, LD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (06) : 1021 - 1027
  • [7] NOISE AND FILTRATION IN MAGNETIC-RESONANCE IMAGING
    MCVEIGH, ER
    HENKELMAN, RM
    BRONSKILL, MJ
    [J]. MEDICAL PHYSICS, 1985, 12 (05) : 586 - 591
  • [8] Measures of performance in nonlinear estimation tasks:: Prediction of estimation performance at low signal-to-noise ratio
    Müller, SP
    Abbey, CK
    Rybicki, FJ
    Moore, SC
    Kijewski, MF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) : 3697 - 3715
  • [9] Cramer-Rao bounds for three-point decomposition of water and fat
    Pineda, AR
    Reeder, SB
    Wen, ZF
    Pelc, NJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (03) : 625 - 635
  • [10] Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): Application with fast spin-echo imaging
    Reeder, SB
    Pineda, AR
    Wen, ZF
    Shimakawa, A
    Yu, HZ
    Brittain, JH
    Gold, GE
    Beaulieu, CH
    Pelc, NJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (03) : 636 - 644