Randomized in vitro and in vivo evaluation of different biopsy needles and devices for breast biopsy

被引:8
作者
Helbich, TH
Rudas, M
Böhm, G
Huber, S
Wagner, T
Taucher, S
Wolf, G
Mostbeck, GH
机构
[1] Univ Vienna, Dept Radiol, AKH, A-1090 Vienna, Austria
[2] Univ Vienna, Dept Clin Pathol, A-1090 Vienna, Austria
[3] Univ Vienna, Dept Special Gynaecol, A-1090 Vienna, Austria
[4] Univ Vienna, Dept Surg, A-1090 Vienna, Austria
[5] Lainz Hosp, Dept Radiol, A-1130 Vienna, Austria
关键词
biopsies; technology; breast neoplasms; biopsy; diagnosis;
D O I
10.1016/S0009-9260(99)91241-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In an experimental study (in vitro and in vivo) we evaluated the efficacy of various biopsy needles/devices for breast biopsy. In vitro, biopsies of five human cadaveric breast specimens were performed using 33 different needles/devices ranging from 14 to 20-gauge. Of these 33 needles/devices, 22 optimally performing needles were selected for the irt vivo study. In the clinical part of the study, 44 breast lesions were randomly biopsied with each of the 22 needles/devices under stereotactic guidance. Tissue specimens were analysed quantitatively and qualitatively. Several automatic long-throw guns (Acecut, Asap, Biopty, Magnum) obtained greater tissue areas and had a better histopathologic score than the conventional type of a side-notch needle like Trucut, an aspiration needle like Surecut, or an end-cut needle like Autovac, The automatic long-throw guns performed better than the short-throw Monopty gun. Regardless of needle size (14-20-gauge), breast biopsies should be routinely performed with automated long-throw side-notch guns (Acecut, Asap, Biopty, Magnum).
引用
收藏
页码:56 / 62
页数:7
相关论文
共 21 条
[1]   CORE BREAST BIOPSY, RESEARCH, AND WHAT NOT TO DO [J].
BURBANK, F ;
BELVILLE, J .
RADIOLOGY, 1992, 185 (03) :639-644
[2]   STEREOTAXIC NEEDLE CORE BIOPSY OF BREAST-LESIONS USING A REGULAR MAMMOGRAPHIC TABLE WITH AN ADAPTABLE STEREOTAXIC DEVICE [J].
CAINES, JS ;
MCPHEE, MD ;
KONOK, GP ;
WRIGHT, BA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (02) :317-321
[3]   NONPALPABLE BREAST-LESIONS - FINDINGS OF STEREOTAXIC NEEDLE-CORE BIOPSY AND FINE-NEEDLE ASPIRATION CYTOLOGY [J].
DOWLATSHAHI, K ;
YAREMKO, ML ;
KLUSKENS, LF ;
JOKICH, PM .
RADIOLOGY, 1991, 181 (03) :745-750
[4]   STEREOTAXIC CORE BIOPSY OF BREAST-LESIONS [J].
DRONKERS, DJ .
RADIOLOGY, 1992, 183 (03) :631-634
[5]   Coaxial technique: Approach to breast core biopsies [J].
Helbich, TH ;
Mayr, W ;
Schick, S ;
Youssefzadeh, S ;
Rudas, M ;
Taucher, S ;
Wagner, T ;
Kelkar, P ;
Wolf, G ;
Thurnher, M ;
Mostbeck, GH .
RADIOLOGY, 1997, 203 (03) :684-690
[6]   AUTOMATED BIOPSY DEVICES - A BLINDED EVALUATION [J].
HOPPER, KD ;
ABENDROTH, CS ;
STURTZ, KW ;
MATTHEWS, YL ;
STEVENS, LA ;
SHIRK, SJ .
RADIOLOGY, 1993, 187 (03) :653-660
[7]   CT PERCUTANEOUS BIOPSY GUNS - COMPARISON OF END-CUT AND SIDE-NOTCH DEVICES IN CADAVERIC SPECIMENS [J].
HOPPER, KD ;
ABENDROTH, CS ;
STURTZ, KW ;
MATTHEWS, YL ;
HARTZEL, JS ;
POTOK, PS .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (01) :195-199
[8]   Large-core biopsy guns: Comparison for yield of breast tissue [J].
Krebs, TL ;
Berg, WA ;
Severson, MJ ;
Magder, LS ;
Goldberg, PA ;
Campassi, C ;
Sun, CCJ .
RADIOLOGY, 1996, 200 (02) :365-368
[9]   RADIOGRAPHY OF MICROCALCIFICATIONS IN STEREOTAXIC MAMMARY CORE BIOPSY SPECIMENS [J].
LIBERMAN, L ;
EVANS, WP ;
DERSHAW, DD ;
HANN, LE ;
DEUTCH, BM ;
ABRAMSON, AF ;
ROSEN, PP .
RADIOLOGY, 1994, 190 (01) :223-225
[10]  
Marchevsky Alberto M., 1994, P125